142
Programming the Enhanced Monitor
Chapter 3
SYSTem:COMMunicate:VXI:ADDRess? <
address
>
SYSTem:COMMunication:VXI:ADDRess
returns the current VXI logical address of
the enhanced monitor device. This address is set by a switch on the back of the
mainframe.
Parameters
Returned Data
Comments
•
The MIN parameter returns the lowest possible address, 1 but does not change
the actual VXI logical address used with the Enhanced Monitor. Similarly, the
MAX parameter returns the largest possible VXI address, 254 but does change
the address used. DEF returns the factory default address of 224.
•
This query will return the VXI logical address unless the mainframe’s power is
off and the enhanced monitor device cannot read it. It then returns the value
32567.
Reset Condition
*RST has no effect on the VXI logical address.
Name
Type
Range of Values
Default
<
address>
enum.
MIN|MAX|DEF
none
Type
Range
Default
Description
int16
1 through 254, or 32767 224
VXI logical address