Programming the Enhanced Monitor
69
Chapter 3
HISTory:POWer[:HISTogram]?
<
supply
>[,MIN|MAX]
HISTory:POWer[:HISTogram]?
<supply> returns the histogram data held for the
<supply>. Ten values are returned for the amount of time the <supply>’s power
spent in ten different power ranges. Those power ranges can be queried with the
MIN
and
MAX
optional parameter. The units of the time values returned are HOURs by
default, but can be changed with the
HISTory:UNIT[:TIME]
command.The units
returned for power ranges queried with the
MIN
or
MAX
parameters are integer
values in milliwatts.
Parameters
Returned Data
Comments
•
HIST:POW? P5
returns 10 time values for +5V supply.
•
HIST:POW? P5,MAX
returns 10 values representing the maximum power for
each power range in milliwatts.
Error Conditions
The following table lists the most common error conditions and causes. Error
numbers and corresponding messages can be found using
SYSTem:ERRor?
query.
Reset Condition
A *RST has no effect on the HISTory subsystem.
Related
Commands
HISTory:POWer:MAXimum?, HISTory:RESet:POWer, HISTory:UNITs[:TIME]
Name
Type
Range
Default
Description
<
supply>
enum. P5, P12, N12, P24,
N24, N5PT2, N2,
TOTal
none
Selects the power supply whose power
histogram is desired. P5 is +5 Vdc
supply, N12 is -12 Vdc supply, etc.
optional
parameter
enum. MIN, MAX
none
Returns the maximum or minimum of
the power range in milliwatts.
Type
Range
Default
Description
<range1>
int32
0-2147483647
none
lowest power range.
<range2>
int32
0-2147483647
none
second power range.
<range3>
int32
0-2147483647
none
third power range.
<range4>
int32
0-2147483647
none
fourth power range.
<range5>
int32
0-2147483647
none
fifth power range.
<range6>
int32
0-2147483647
none
sixth power range.
<range7>
int32
0-2147483647
none
seventh power range.
<range8>
int32
0-2147483647
none
eighth power range.
<range9>
int32
0-2147483647
none
ninth power range.
<range10>
int32
0-2147483647
none
highest power range.
Number
Message
Probable Cause
-224
“Illegal Parameter”
The
<supply> or optional parameter not correct.
-109
“Missing Parameter”
The
<supply> parameter was omitted.
-311
“Memory error”
The historical data’s memory is not available.
-221
“Settings Conflict”
The power supply doesn’t support power
measurements.
Содержание E8402A
Страница 8: ...10 Contents ...
Страница 12: ...14 Notes ...
Страница 14: ......
Страница 26: ...28 Getting Started Chapter 1 ...
Страница 42: ...44 Using the Enhanced Monitor Chapter 2 ...
Страница 214: ...216 Servicing Your Mainframe Chapter 5 ...
Страница 226: ...228 HP E8402 E8404A Product Specifications Appendix A ...