36
2
Operating the MSD
To view MSD analyzer temperature and vacuum status
To view MSD analyzer temperature and vacuum status
Software changes
The software is revised periodically. If the steps in this procedure do not match
your MSD ChemStation software, refer to the manuals and online help supplied
with the software for more information.
1
In Instrument Control view, select
Edit MS Tune Parameters
from the
Instrument menu.
2
Select the tune file you plan to use with your method from the Load MS
Tune File dialog box.
3
Analyzer temperatures and vacuum status are displayed in the
Zones
field.
Unless you have just begun the pumpdown process, the foreline pressure should
be less than 300 mTorr, or the turbo pump should be running at least 80% speed.
MSD heaters remain off as long as the diffusion pump is cold or the turbo pump is
operating at less than 80%. Normally, the foreline pressure will be below 100
mTorr, or the turbo pump speed will be at 100%.
The MSD heaters turn off at the beginning of the vent cycle, and turn on at the end
of the pumpdown cycle. Note that the reported setpoints will not change during
venting or pumpdown, even though both the MSD zones are turned off.
Содержание 5973
Страница 1: ...H Hardware Manual HP 5973 Mass Selective Detector ...
Страница 3: ...HP 5973 Mass Selective Detector Hardware Manual ...
Страница 25: ...1 Installing GC Columns To condition a capillary column 25 ...
Страница 30: ...30 ...
Страница 37: ...2 Operating the MSD To view MSD analyzer temperature and vacuum status 37 ...
Страница 39: ...2 Operating the MSD To set monitors for MSD temperature and vacuum status 39 ...
Страница 41: ...2 Operating the MSD To set the MSD analyzer temperatures 41 ...
Страница 43: ...2 Operating the MSD To set the interface temperature from the ChemStation 43 ...
Страница 51: ...2 Operating the MSD To remove the MSD covers 51 Upper cover Lower cover ...
Страница 92: ...92 ...
Страница 103: ...4 Maintaining the MSD To check and add foreline pump oil 103 Lower line Oil level window Fill cap Upper line ...
Страница 105: ...4 Maintaining the MSD To drain the foreline pump 105 Fill cap Drain plug Pump motor ...
Страница 107: ...4 Maintaining the MSD To refill the foreline pump 107 Fill cap Upper line Oil level window Drain plug ...
Страница 109: ...4 Maintaining the MSD To replace the oil trap 109 Oil trap Clamp O ring assembly Oil trap adapter Foreline pump ...
Страница 111: ...4 Maintaining the MSD To check the diffusion pump fluid 111 Fan Fluid level window sight glass ...
Страница 113: ...4 Maintaining the MSD To separate the MSD from the GC 113 ...
Страница 121: ...4 Maintaining the MSD To reconnect the MSD to the GC 121 ...
Страница 123: ...4 Maintaining the MSD To remove the calibration vial 123 Calibration valve Collar Calibration vial ...
Страница 125: ...4 Maintaining the MSD To refill and reinstall the calibration vial 125 Calibration valve Collar Calibration vial 6 mm ...
Страница 141: ...4 Maintaining the MSD To lubricate the side plate O ring 141 Side plate O ring ...
Страница 159: ...4 Maintaining the MSD To remove a filament 159 Filament 2 Filament 1 ...
Страница 161: ...4 Maintaining the MSD To reinstall a filament 161 Filament 2 Filament 1 ...
Страница 210: ...210 ...
Страница 211: ...6 GC MSD Interface This chapter describes the function of the GC MSD interface GC MSD interface 212 ...
Страница 214: ...214 ...
Страница 232: ...232 7 Analyzer Analyzer heaters and radiators ...
Страница 280: ...280 ...
Страница 294: ...294 Index ...
Страница 296: ...294 Chapter ...
Страница 297: ...H Manual Part Number G1099 90027 Copyright 1998 Hewlett Packard Printed in USA 11 98 ...