Troubleshooting the MSD
Error messages
The system is in vent state
Although the message says the system is in vent state, if the fault has just occurred
the MSD may actually still be under vacuum and the diffusion pump may still be
hot or the turbo pump may still be at high speed. Wait at least 30 minutes after
seeing this message before you actually vent the MSD.
•
System was vented on purpose (no problem)
•
Fan fault has turned off the high vacuum pump (power cycle the MSD to clear
the fault)
•
Fuse for the high vacuum pump has failed
•
Diffusion pump heater cartridge has failed
•
MSD electronics are not working correctly
There is no emission current
•
Filament is not connected properly
•
Filament has failed
•
MSD electronics are not working correctly
There is not enough signal to begin tune
•
Poor mass axis calibration
•
Amu gain or offset is too high
•
Calibration vial is empty or almost empty
•
Excessive pressure in the vacuum manifold
•
Air leak
•
Electron multiplier voltage is too low
•
Signal cable is not connected
•
Electrical leads to the detector are not connected correctly
•
Electrical leads to the ion source are not connected correctly
•
Filament shorted to the source body
& $ 8 7 , 2 1
Venting the MSD too soon after this message appears can result in the backstreaming of
a diffusion pump or damage to a turbomolecular pump.
Содержание 5973
Страница 1: ...H Hardware Manual HP 5973 Mass Selective Detector ...
Страница 3: ...HP 5973 Mass Selective Detector Hardware Manual ...
Страница 25: ...1 Installing GC Columns To condition a capillary column 25 ...
Страница 30: ...30 ...
Страница 37: ...2 Operating the MSD To view MSD analyzer temperature and vacuum status 37 ...
Страница 39: ...2 Operating the MSD To set monitors for MSD temperature and vacuum status 39 ...
Страница 41: ...2 Operating the MSD To set the MSD analyzer temperatures 41 ...
Страница 43: ...2 Operating the MSD To set the interface temperature from the ChemStation 43 ...
Страница 51: ...2 Operating the MSD To remove the MSD covers 51 Upper cover Lower cover ...
Страница 92: ...92 ...
Страница 103: ...4 Maintaining the MSD To check and add foreline pump oil 103 Lower line Oil level window Fill cap Upper line ...
Страница 105: ...4 Maintaining the MSD To drain the foreline pump 105 Fill cap Drain plug Pump motor ...
Страница 107: ...4 Maintaining the MSD To refill the foreline pump 107 Fill cap Upper line Oil level window Drain plug ...
Страница 109: ...4 Maintaining the MSD To replace the oil trap 109 Oil trap Clamp O ring assembly Oil trap adapter Foreline pump ...
Страница 111: ...4 Maintaining the MSD To check the diffusion pump fluid 111 Fan Fluid level window sight glass ...
Страница 113: ...4 Maintaining the MSD To separate the MSD from the GC 113 ...
Страница 121: ...4 Maintaining the MSD To reconnect the MSD to the GC 121 ...
Страница 123: ...4 Maintaining the MSD To remove the calibration vial 123 Calibration valve Collar Calibration vial ...
Страница 125: ...4 Maintaining the MSD To refill and reinstall the calibration vial 125 Calibration valve Collar Calibration vial 6 mm ...
Страница 141: ...4 Maintaining the MSD To lubricate the side plate O ring 141 Side plate O ring ...
Страница 159: ...4 Maintaining the MSD To remove a filament 159 Filament 2 Filament 1 ...
Страница 161: ...4 Maintaining the MSD To reinstall a filament 161 Filament 2 Filament 1 ...
Страница 210: ...210 ...
Страница 211: ...6 GC MSD Interface This chapter describes the function of the GC MSD interface GC MSD interface 212 ...
Страница 214: ...214 ...
Страница 232: ...232 7 Analyzer Analyzer heaters and radiators ...
Страница 280: ...280 ...
Страница 294: ...294 Index ...
Страница 296: ...294 Chapter ...
Страница 297: ...H Manual Part Number G1099 90027 Copyright 1998 Hewlett Packard Printed in USA 11 98 ...