Operating the MSD
To vent the MSD
To vent the MSD
Software changes
The software is revised periodically. If the steps in this procedure do not match
your MSD ChemStation software, refer to the manuals and online help supplied
with the software for more information.
If the data system is not working correctly, the MSD can still be vented safely. See
page 64.
1
If your system is equipped with a gauge controller, switch off the triode
gauge controller.
2
Select
Diagnostics/Vacuum Control
from the View menu.
3
Select
Vent
from the Vacuum menu. Follow the instructions presented.
The vent program cools the GC/MSD interface heater and the GC oven. The pro-
gram turns off the analyzer heaters and the high vacuum pump. The software will
prompt you when it is safe to switch off the power.
4
When prompted, turn off the MSD power switch.
5
Unplug the MSD power cord.
6
Remove the upper MSD cover (page 50).
7
Turn the vent valve knob counterclockwise
only
1/4 to 3/4 turns to admit
air into the vacuum manifold.
Do
not
turn the knob too far. Be sure to retighten the knob before pumping down.
: $ 5 1 , 1 *
,I\RXDUHXVLQJK\GURJHQDVDFDUULHUJDVWKHFDUULHUJDVIORZPXVWEHRIIEHIRUHWXUQLQJ
RIIWKH06'SRZHU,IWKHIRUHOLQHSXPSLVRIIK\GURJHQZLOODFFXPXODWHLQWKH06'DQGDQ
H[SORVLRQPD\RFFXU5HDGWKH
+\GURJHQ&DUULHU*DV6DIHW\*XLGH
EHIRUH
RSHUDWLQJWKH06'ZLWKK\GURJHQFDUULHUJDV
& $ 8 7 , 2 1
Be sure the GC oven and the GC/MSD interface are cool before turning off carrier gas flow.
Содержание 5973
Страница 1: ...H Hardware Manual HP 5973 Mass Selective Detector ...
Страница 3: ...HP 5973 Mass Selective Detector Hardware Manual ...
Страница 25: ...1 Installing GC Columns To condition a capillary column 25 ...
Страница 30: ...30 ...
Страница 37: ...2 Operating the MSD To view MSD analyzer temperature and vacuum status 37 ...
Страница 39: ...2 Operating the MSD To set monitors for MSD temperature and vacuum status 39 ...
Страница 41: ...2 Operating the MSD To set the MSD analyzer temperatures 41 ...
Страница 43: ...2 Operating the MSD To set the interface temperature from the ChemStation 43 ...
Страница 51: ...2 Operating the MSD To remove the MSD covers 51 Upper cover Lower cover ...
Страница 92: ...92 ...
Страница 103: ...4 Maintaining the MSD To check and add foreline pump oil 103 Lower line Oil level window Fill cap Upper line ...
Страница 105: ...4 Maintaining the MSD To drain the foreline pump 105 Fill cap Drain plug Pump motor ...
Страница 107: ...4 Maintaining the MSD To refill the foreline pump 107 Fill cap Upper line Oil level window Drain plug ...
Страница 109: ...4 Maintaining the MSD To replace the oil trap 109 Oil trap Clamp O ring assembly Oil trap adapter Foreline pump ...
Страница 111: ...4 Maintaining the MSD To check the diffusion pump fluid 111 Fan Fluid level window sight glass ...
Страница 113: ...4 Maintaining the MSD To separate the MSD from the GC 113 ...
Страница 121: ...4 Maintaining the MSD To reconnect the MSD to the GC 121 ...
Страница 123: ...4 Maintaining the MSD To remove the calibration vial 123 Calibration valve Collar Calibration vial ...
Страница 125: ...4 Maintaining the MSD To refill and reinstall the calibration vial 125 Calibration valve Collar Calibration vial 6 mm ...
Страница 141: ...4 Maintaining the MSD To lubricate the side plate O ring 141 Side plate O ring ...
Страница 159: ...4 Maintaining the MSD To remove a filament 159 Filament 2 Filament 1 ...
Страница 161: ...4 Maintaining the MSD To reinstall a filament 161 Filament 2 Filament 1 ...
Страница 210: ...210 ...
Страница 211: ...6 GC MSD Interface This chapter describes the function of the GC MSD interface GC MSD interface 212 ...
Страница 214: ...214 ...
Страница 232: ...232 7 Analyzer Analyzer heaters and radiators ...
Страница 280: ...280 ...
Страница 294: ...294 Index ...
Страница 296: ...294 Chapter ...
Страница 297: ...H Manual Part Number G1099 90027 Copyright 1998 Hewlett Packard Printed in USA 11 98 ...