4
Maintaining the MSD
To remove the GC/MSD interface heater and sensor
173
To remove the GC/MSD interface heater and sensor
Materials needed:
Screwdriver, TORX T-15 (8710-1622)
1
Vent the MSD (page 52).
Make sure you turn off the GC/MSD interface heater. This heater is controlled and
powered by the GC.
2
Separate the MSD from the GC (page 112).
3
Remove the cover from the GC/MSD interface.
4
Slide the insulation off of the GC/MSD interface.
5
Loosen the two heater sleeve screws.
6
Slide the heater sleeve off of the GC/MSD interface.
It may be necessary to
gently
pry open the slot in the heater sleeve to loosen the
heater sleeve from the interface.
7
Loosen the setscrew and remove the heater and temperature sensor from
the heater sleeve.
Heat and oxidation often result in a heater, or less frequently a temperature sen-
sor, being “welded” inside the heater sleeve. The holes for the heater and sensor
pass all the way through the heater sleeve. A rod can be inserted to drive the
stuck part out. However, to function correctly the heater and sensor must have
perfect contact with their holes. If a heater or sensor is difficult to remove, the
holes will probably be damaged enough that the heater sleeve should be replaced.
Polishing the holes is not an acceptable solution since it will enlarge the holes.
: $ 5 1 , 1 *
7KH*&06'LQWHUIDFHRSHUDWHVDYHU\KLJKWHPSHUDWXUHV,WLVDOVRZHOOLQVXODWHG0DNH
VXUHWKHLQWHUIDFHLVFRRO
EHIRUH
\RXWRXFKLW
& $ 8 7 , 2 1
Installing a new heater and sensor in a damaged heater sleeve will result in poor
performance of the heated zone and could reduce the lifetime of the new parts.
Содержание 5973
Страница 1: ...H Hardware Manual HP 5973 Mass Selective Detector ...
Страница 3: ...HP 5973 Mass Selective Detector Hardware Manual ...
Страница 25: ...1 Installing GC Columns To condition a capillary column 25 ...
Страница 30: ...30 ...
Страница 37: ...2 Operating the MSD To view MSD analyzer temperature and vacuum status 37 ...
Страница 39: ...2 Operating the MSD To set monitors for MSD temperature and vacuum status 39 ...
Страница 41: ...2 Operating the MSD To set the MSD analyzer temperatures 41 ...
Страница 43: ...2 Operating the MSD To set the interface temperature from the ChemStation 43 ...
Страница 51: ...2 Operating the MSD To remove the MSD covers 51 Upper cover Lower cover ...
Страница 92: ...92 ...
Страница 103: ...4 Maintaining the MSD To check and add foreline pump oil 103 Lower line Oil level window Fill cap Upper line ...
Страница 105: ...4 Maintaining the MSD To drain the foreline pump 105 Fill cap Drain plug Pump motor ...
Страница 107: ...4 Maintaining the MSD To refill the foreline pump 107 Fill cap Upper line Oil level window Drain plug ...
Страница 109: ...4 Maintaining the MSD To replace the oil trap 109 Oil trap Clamp O ring assembly Oil trap adapter Foreline pump ...
Страница 111: ...4 Maintaining the MSD To check the diffusion pump fluid 111 Fan Fluid level window sight glass ...
Страница 113: ...4 Maintaining the MSD To separate the MSD from the GC 113 ...
Страница 121: ...4 Maintaining the MSD To reconnect the MSD to the GC 121 ...
Страница 123: ...4 Maintaining the MSD To remove the calibration vial 123 Calibration valve Collar Calibration vial ...
Страница 125: ...4 Maintaining the MSD To refill and reinstall the calibration vial 125 Calibration valve Collar Calibration vial 6 mm ...
Страница 141: ...4 Maintaining the MSD To lubricate the side plate O ring 141 Side plate O ring ...
Страница 159: ...4 Maintaining the MSD To remove a filament 159 Filament 2 Filament 1 ...
Страница 161: ...4 Maintaining the MSD To reinstall a filament 161 Filament 2 Filament 1 ...
Страница 210: ...210 ...
Страница 211: ...6 GC MSD Interface This chapter describes the function of the GC MSD interface GC MSD interface 212 ...
Страница 214: ...214 ...
Страница 232: ...232 7 Analyzer Analyzer heaters and radiators ...
Страница 280: ...280 ...
Страница 294: ...294 Index ...
Страница 296: ...294 Chapter ...
Страница 297: ...H Manual Part Number G1099 90027 Copyright 1998 Hewlett Packard Printed in USA 11 98 ...