56
2
Operating the MSD
To close the vacuum manifold
To close the vacuum manifold
Materials needed:
Gloves, clean, lint-free
large (8650-0030)
small (8650-0029)
1
Make sure all the internal analyzer electrical leads are correctly attached.
2
Check the side plate O-ring.
Make sure the O-ring has a
very
light coat of Apiezon high vacuum grease. If the
O-ring is very dry, it may not seal well. If the O-ring looks shiny, it has too much
grease on it. See page 140 for instructions for lubricating the side plate O-ring.
3
Close the side plate.
4
Reconnect the side board control cable and source power cable to the side
board.
5
Make sure the vent valve is closed.
6
Pump down the MSD (page 58).
7
Gently
hand tighten the front side plate thumbscrew
This is only necessary for CI MSDs, or if hydrogen or other flammable or toxic sub-
stance is used for carrier gas.
8
Once the MSD has pumped down, reinstall the upper MSD cover.
Wait until after pumpdown to reinstall the upper MSD cover.
: $ 5 1 , 1 *
7KLVWKXPEVFUHZPXVWEHIDVWHQHGIRU&,RSHUDWLRQRULIK\GURJHQRURWKHUKD]DUGRXVJDV
LVEHLQJXVHGDVWKH*&FDUULHUJDV,QWKHXQOLNHO\HYHQWRIDQH[SORVLRQLWPD\SUHYHQWWKH
VLGHSODWHIURPRSHQLQJ
& $ 8 7 , 2 1
Do not overtighten the thumbscrew; it can cause air leaks or prevent successful
pumpdown. Do not use a screwdriver to tighten the thumbscrew.
Содержание 5973
Страница 1: ...H Hardware Manual HP 5973 Mass Selective Detector ...
Страница 3: ...HP 5973 Mass Selective Detector Hardware Manual ...
Страница 25: ...1 Installing GC Columns To condition a capillary column 25 ...
Страница 30: ...30 ...
Страница 37: ...2 Operating the MSD To view MSD analyzer temperature and vacuum status 37 ...
Страница 39: ...2 Operating the MSD To set monitors for MSD temperature and vacuum status 39 ...
Страница 41: ...2 Operating the MSD To set the MSD analyzer temperatures 41 ...
Страница 43: ...2 Operating the MSD To set the interface temperature from the ChemStation 43 ...
Страница 51: ...2 Operating the MSD To remove the MSD covers 51 Upper cover Lower cover ...
Страница 92: ...92 ...
Страница 103: ...4 Maintaining the MSD To check and add foreline pump oil 103 Lower line Oil level window Fill cap Upper line ...
Страница 105: ...4 Maintaining the MSD To drain the foreline pump 105 Fill cap Drain plug Pump motor ...
Страница 107: ...4 Maintaining the MSD To refill the foreline pump 107 Fill cap Upper line Oil level window Drain plug ...
Страница 109: ...4 Maintaining the MSD To replace the oil trap 109 Oil trap Clamp O ring assembly Oil trap adapter Foreline pump ...
Страница 111: ...4 Maintaining the MSD To check the diffusion pump fluid 111 Fan Fluid level window sight glass ...
Страница 113: ...4 Maintaining the MSD To separate the MSD from the GC 113 ...
Страница 121: ...4 Maintaining the MSD To reconnect the MSD to the GC 121 ...
Страница 123: ...4 Maintaining the MSD To remove the calibration vial 123 Calibration valve Collar Calibration vial ...
Страница 125: ...4 Maintaining the MSD To refill and reinstall the calibration vial 125 Calibration valve Collar Calibration vial 6 mm ...
Страница 141: ...4 Maintaining the MSD To lubricate the side plate O ring 141 Side plate O ring ...
Страница 159: ...4 Maintaining the MSD To remove a filament 159 Filament 2 Filament 1 ...
Страница 161: ...4 Maintaining the MSD To reinstall a filament 161 Filament 2 Filament 1 ...
Страница 210: ...210 ...
Страница 211: ...6 GC MSD Interface This chapter describes the function of the GC MSD interface GC MSD interface 212 ...
Страница 214: ...214 ...
Страница 232: ...232 7 Analyzer Analyzer heaters and radiators ...
Страница 280: ...280 ...
Страница 294: ...294 Index ...
Страница 296: ...294 Chapter ...
Страница 297: ...H Manual Part Number G1099 90027 Copyright 1998 Hewlett Packard Printed in USA 11 98 ...