Troubleshooting the MSD
Mass spectral symptoms
75
Mass spectral symptoms
This section describes symptoms you might observe in mass spectra. Some of
these symptoms will appear in the mass spectra of samples. Others you will
observe only in a tune report. Some of these symptoms have causes that can be
corrected by the operator. Others, however, require service by a Hewlett-Packard
service representative.
Two symptoms listed under Chromatographic symptoms:
If sensitivity is poor
and
If repeatability is poor
, also apply to mass spectra.
Isotopes are missing or isotope ratios are incorrect
•
Peaks are too wide or too narrow
•
Scan speed is too high (scan mode)
•
Dwell time is too short (SIM mode)
•
Electron multiplier voltage is too high
•
Repeller voltage is too high
•
High background
•
Dirty ion source
High background
•
Pressure in the vacuum manifold is too high
•
Air leak
•
Contamination
High abundances at
m/z
18, 28, 32, and 44 or at
m/z
14 and 16
•
System was recently vented (residual air and water)
•
Air leak
Large peaks at
m/z
14 and 16 are symptomatic of especially large leaks.
Содержание 5973
Страница 1: ...H Hardware Manual HP 5973 Mass Selective Detector ...
Страница 3: ...HP 5973 Mass Selective Detector Hardware Manual ...
Страница 25: ...1 Installing GC Columns To condition a capillary column 25 ...
Страница 30: ...30 ...
Страница 37: ...2 Operating the MSD To view MSD analyzer temperature and vacuum status 37 ...
Страница 39: ...2 Operating the MSD To set monitors for MSD temperature and vacuum status 39 ...
Страница 41: ...2 Operating the MSD To set the MSD analyzer temperatures 41 ...
Страница 43: ...2 Operating the MSD To set the interface temperature from the ChemStation 43 ...
Страница 51: ...2 Operating the MSD To remove the MSD covers 51 Upper cover Lower cover ...
Страница 92: ...92 ...
Страница 103: ...4 Maintaining the MSD To check and add foreline pump oil 103 Lower line Oil level window Fill cap Upper line ...
Страница 105: ...4 Maintaining the MSD To drain the foreline pump 105 Fill cap Drain plug Pump motor ...
Страница 107: ...4 Maintaining the MSD To refill the foreline pump 107 Fill cap Upper line Oil level window Drain plug ...
Страница 109: ...4 Maintaining the MSD To replace the oil trap 109 Oil trap Clamp O ring assembly Oil trap adapter Foreline pump ...
Страница 111: ...4 Maintaining the MSD To check the diffusion pump fluid 111 Fan Fluid level window sight glass ...
Страница 113: ...4 Maintaining the MSD To separate the MSD from the GC 113 ...
Страница 121: ...4 Maintaining the MSD To reconnect the MSD to the GC 121 ...
Страница 123: ...4 Maintaining the MSD To remove the calibration vial 123 Calibration valve Collar Calibration vial ...
Страница 125: ...4 Maintaining the MSD To refill and reinstall the calibration vial 125 Calibration valve Collar Calibration vial 6 mm ...
Страница 141: ...4 Maintaining the MSD To lubricate the side plate O ring 141 Side plate O ring ...
Страница 159: ...4 Maintaining the MSD To remove a filament 159 Filament 2 Filament 1 ...
Страница 161: ...4 Maintaining the MSD To reinstall a filament 161 Filament 2 Filament 1 ...
Страница 210: ...210 ...
Страница 211: ...6 GC MSD Interface This chapter describes the function of the GC MSD interface GC MSD interface 212 ...
Страница 214: ...214 ...
Страница 232: ...232 7 Analyzer Analyzer heaters and radiators ...
Страница 280: ...280 ...
Страница 294: ...294 Index ...
Страница 296: ...294 Chapter ...
Страница 297: ...H Manual Part Number G1099 90027 Copyright 1998 Hewlett Packard Printed in USA 11 98 ...