250
8
Electronics
Status display and power switch
Status display and power switch
The MSD has a status display on the lower left of the front of the instrument. It
includes three status LEDs.
Each LED provides information about the status of the MSD. In addition, if power
to the MSD is interrupted, all three LEDs will blink until the data system reestab-
lishes contact with the MSD and downloads the proper information to the HP-IB/
MS control card.
Power status LED
The power status LED lights when the MSD is connected to a power source and
the power switch is on.
Analyzer status LED
The analyzer (ion source, mass filter, and detector) is turned off except when the
MSD is acquiring data or tuning. This helps extend the life spans of the filaments
and electron multiplier horn. The analyzer status LED lights when the MSD
receives the MS ON signal from the data system. The analyzer LED should always
be on during data acquisition or tuning.
Cal valve status LED
The cal (calibration) valve status LED lights when the data system instructs the
calibration valve to open.
Power switch
The power switch is part of the electronics module, and is located on the lower left
of the front of the MSD. It is used to switch the MSD and foreline pump on and off.
& $ 8 7 , 2 1
Do not switch the MSD off unless it has completed the vent program. Incorrect shutdown
can seriously damage the MSD.
Содержание 5973
Страница 1: ...H Hardware Manual HP 5973 Mass Selective Detector ...
Страница 3: ...HP 5973 Mass Selective Detector Hardware Manual ...
Страница 25: ...1 Installing GC Columns To condition a capillary column 25 ...
Страница 30: ...30 ...
Страница 37: ...2 Operating the MSD To view MSD analyzer temperature and vacuum status 37 ...
Страница 39: ...2 Operating the MSD To set monitors for MSD temperature and vacuum status 39 ...
Страница 41: ...2 Operating the MSD To set the MSD analyzer temperatures 41 ...
Страница 43: ...2 Operating the MSD To set the interface temperature from the ChemStation 43 ...
Страница 51: ...2 Operating the MSD To remove the MSD covers 51 Upper cover Lower cover ...
Страница 92: ...92 ...
Страница 103: ...4 Maintaining the MSD To check and add foreline pump oil 103 Lower line Oil level window Fill cap Upper line ...
Страница 105: ...4 Maintaining the MSD To drain the foreline pump 105 Fill cap Drain plug Pump motor ...
Страница 107: ...4 Maintaining the MSD To refill the foreline pump 107 Fill cap Upper line Oil level window Drain plug ...
Страница 109: ...4 Maintaining the MSD To replace the oil trap 109 Oil trap Clamp O ring assembly Oil trap adapter Foreline pump ...
Страница 111: ...4 Maintaining the MSD To check the diffusion pump fluid 111 Fan Fluid level window sight glass ...
Страница 113: ...4 Maintaining the MSD To separate the MSD from the GC 113 ...
Страница 121: ...4 Maintaining the MSD To reconnect the MSD to the GC 121 ...
Страница 123: ...4 Maintaining the MSD To remove the calibration vial 123 Calibration valve Collar Calibration vial ...
Страница 125: ...4 Maintaining the MSD To refill and reinstall the calibration vial 125 Calibration valve Collar Calibration vial 6 mm ...
Страница 141: ...4 Maintaining the MSD To lubricate the side plate O ring 141 Side plate O ring ...
Страница 159: ...4 Maintaining the MSD To remove a filament 159 Filament 2 Filament 1 ...
Страница 161: ...4 Maintaining the MSD To reinstall a filament 161 Filament 2 Filament 1 ...
Страница 210: ...210 ...
Страница 211: ...6 GC MSD Interface This chapter describes the function of the GC MSD interface GC MSD interface 212 ...
Страница 214: ...214 ...
Страница 232: ...232 7 Analyzer Analyzer heaters and radiators ...
Страница 280: ...280 ...
Страница 294: ...294 Index ...
Страница 296: ...294 Chapter ...
Страница 297: ...H Manual Part Number G1099 90027 Copyright 1998 Hewlett Packard Printed in USA 11 98 ...