194
5
Vacuum System
Foreline pump
Foreline pump
The foreline pump reduces the pressure in the vacuum manifold so the high vac-
uum pump can operate It also pumps away the gas load from the high vacuum
pump. The foreline pump is connected to the high vacuum pump by a 130-cm hose
called the foreline hose.
The foreline pump is a two-stage rotary-vane pump. The foreline pump turns on
when the MSD power is turned on. The foreline pump has a built-in anti-suckback
valve to help prevent backstreaming in the event of a power failure.
The foreline pump can be placed under the vacuum manifold at the rear of the
MSD (with the exhaust outlet to the rear), or on the floor below the MSD.
The MSD shipping kit includes an oil trap (not shown) that can be used to filter
pump oil out of the foreline pump exhaust. This trap stops
only
pump oil. If you
are analyzing toxic chemicals or using toxic solvents, do not use the trap. Instead,
install a hose to take the foreline pump exhaust outside or to a fume hood.
A window (sight glass) in the front of the foreline pump shows the level of the
foreline pump oil. There are two marks next to the window. The level of the pump
oil should never be above the upper mark or below the lower mark. If the level of
pump oil is near the lower mark, add foreline pump oil.
See Also
To check and add foreline pump oil
, page 102
: $ 5 1 , 1 *
7KHIRUHOLQHSXPSH[KDXVWFRQWDLQVWUDFHVRIVROYHQWVDQDO\WHVDQGIRUHOLQHSXPSRLO7KH
VXSSOLHGRLOWUDSVWRSVRQO\SXPSRLO,WGRHVQRWWUDSRUILOWHURXWWR[LFFKHPLFDOV,I\RXDUH
XVLQJWR[LFVROYHQWVRUDQDO\]LQJWR[LFFKHPLFDOVUHPRYHWKHRLOWUDSDQGLQVWDOODKRVHWR
WDNHWKHIRUHOLQHSXPSH[KDXVWRXWVLGHRUWRDIXPHKRRG
& $ 8 7 , 2 1
Do not place the foreline pump near any equipment that is sensitive to vibration.
& $ 8 7 , 2 1
The ballast control knob controls the amount of air allowed into the pump. Keep the ballast
control closed (fully clockwise) at all times.
Содержание 5973
Страница 1: ...H Hardware Manual HP 5973 Mass Selective Detector ...
Страница 3: ...HP 5973 Mass Selective Detector Hardware Manual ...
Страница 25: ...1 Installing GC Columns To condition a capillary column 25 ...
Страница 30: ...30 ...
Страница 37: ...2 Operating the MSD To view MSD analyzer temperature and vacuum status 37 ...
Страница 39: ...2 Operating the MSD To set monitors for MSD temperature and vacuum status 39 ...
Страница 41: ...2 Operating the MSD To set the MSD analyzer temperatures 41 ...
Страница 43: ...2 Operating the MSD To set the interface temperature from the ChemStation 43 ...
Страница 51: ...2 Operating the MSD To remove the MSD covers 51 Upper cover Lower cover ...
Страница 92: ...92 ...
Страница 103: ...4 Maintaining the MSD To check and add foreline pump oil 103 Lower line Oil level window Fill cap Upper line ...
Страница 105: ...4 Maintaining the MSD To drain the foreline pump 105 Fill cap Drain plug Pump motor ...
Страница 107: ...4 Maintaining the MSD To refill the foreline pump 107 Fill cap Upper line Oil level window Drain plug ...
Страница 109: ...4 Maintaining the MSD To replace the oil trap 109 Oil trap Clamp O ring assembly Oil trap adapter Foreline pump ...
Страница 111: ...4 Maintaining the MSD To check the diffusion pump fluid 111 Fan Fluid level window sight glass ...
Страница 113: ...4 Maintaining the MSD To separate the MSD from the GC 113 ...
Страница 121: ...4 Maintaining the MSD To reconnect the MSD to the GC 121 ...
Страница 123: ...4 Maintaining the MSD To remove the calibration vial 123 Calibration valve Collar Calibration vial ...
Страница 125: ...4 Maintaining the MSD To refill and reinstall the calibration vial 125 Calibration valve Collar Calibration vial 6 mm ...
Страница 141: ...4 Maintaining the MSD To lubricate the side plate O ring 141 Side plate O ring ...
Страница 159: ...4 Maintaining the MSD To remove a filament 159 Filament 2 Filament 1 ...
Страница 161: ...4 Maintaining the MSD To reinstall a filament 161 Filament 2 Filament 1 ...
Страница 210: ...210 ...
Страница 211: ...6 GC MSD Interface This chapter describes the function of the GC MSD interface GC MSD interface 212 ...
Страница 214: ...214 ...
Страница 232: ...232 7 Analyzer Analyzer heaters and radiators ...
Страница 280: ...280 ...
Страница 294: ...294 Index ...
Страница 296: ...294 Chapter ...
Страница 297: ...H Manual Part Number G1099 90027 Copyright 1998 Hewlett Packard Printed in USA 11 98 ...