4
Maintaining the MSD
To replace the primary fuses
181
To replace the primary fuses
Materials needed:
Fuse, T8 A, 250 V (2110-0969) – 2 required
Screwdriver, flat-blade (8730-0002)
The most likely cause of failure of the primary fuses is a problem with the foreline
pump. If the primary fuses in your MSD fail, check the foreline pump.
1
Switch off the MSD and unplug the power cord from the electrical outlet.
If one of the primary fuses has failed, the MSD will already be off, but for safety
you should switch off the MSD and unplug the power cord.
2
Vent the vacuum manifold by opening the vent valve knob.
3
Turn one of the fuse holders counterclockwise until it pops out.
The fuse holders are spring loaded.
4
Remove the old fuse from the fuse holder.
5
Install a new fuse in the fuse holder.
6
Reinstall the fuse holder.
: $ 5 1 , 1 *
1HYHUUHSODFHWKHSULPDU\IXVHVZKLOHWKH06'LVFRQQHFWHGWRDSRZHUVRXUFH
& $ 8 7 , 2 1
If a fuse failure turns off the MSD, wait at least 30 minutes before venting the instrument.
Venting sooner could cause a diffusion pump to backstream or could damage a
turbomolecular pump.
: $ 5 1 , 1 *
,I\RXDUHXVLQJK\GURJHQDVD*&FDUULHUJDVDSRZHUIDLOXUHPD\DOORZLWWRDFFXPXODWHLQ
WKHYDFXXPPDQLIROG,QWKDWFDVHIXUWKHUSUHFDXWLRQVDUHUHTXLUHG6HHWKH
+\GURJHQ
&DUULHU*DV6DIHW\*XLGH
Содержание 5973
Страница 1: ...H Hardware Manual HP 5973 Mass Selective Detector ...
Страница 3: ...HP 5973 Mass Selective Detector Hardware Manual ...
Страница 25: ...1 Installing GC Columns To condition a capillary column 25 ...
Страница 30: ...30 ...
Страница 37: ...2 Operating the MSD To view MSD analyzer temperature and vacuum status 37 ...
Страница 39: ...2 Operating the MSD To set monitors for MSD temperature and vacuum status 39 ...
Страница 41: ...2 Operating the MSD To set the MSD analyzer temperatures 41 ...
Страница 43: ...2 Operating the MSD To set the interface temperature from the ChemStation 43 ...
Страница 51: ...2 Operating the MSD To remove the MSD covers 51 Upper cover Lower cover ...
Страница 92: ...92 ...
Страница 103: ...4 Maintaining the MSD To check and add foreline pump oil 103 Lower line Oil level window Fill cap Upper line ...
Страница 105: ...4 Maintaining the MSD To drain the foreline pump 105 Fill cap Drain plug Pump motor ...
Страница 107: ...4 Maintaining the MSD To refill the foreline pump 107 Fill cap Upper line Oil level window Drain plug ...
Страница 109: ...4 Maintaining the MSD To replace the oil trap 109 Oil trap Clamp O ring assembly Oil trap adapter Foreline pump ...
Страница 111: ...4 Maintaining the MSD To check the diffusion pump fluid 111 Fan Fluid level window sight glass ...
Страница 113: ...4 Maintaining the MSD To separate the MSD from the GC 113 ...
Страница 121: ...4 Maintaining the MSD To reconnect the MSD to the GC 121 ...
Страница 123: ...4 Maintaining the MSD To remove the calibration vial 123 Calibration valve Collar Calibration vial ...
Страница 125: ...4 Maintaining the MSD To refill and reinstall the calibration vial 125 Calibration valve Collar Calibration vial 6 mm ...
Страница 141: ...4 Maintaining the MSD To lubricate the side plate O ring 141 Side plate O ring ...
Страница 159: ...4 Maintaining the MSD To remove a filament 159 Filament 2 Filament 1 ...
Страница 161: ...4 Maintaining the MSD To reinstall a filament 161 Filament 2 Filament 1 ...
Страница 210: ...210 ...
Страница 211: ...6 GC MSD Interface This chapter describes the function of the GC MSD interface GC MSD interface 212 ...
Страница 214: ...214 ...
Страница 232: ...232 7 Analyzer Analyzer heaters and radiators ...
Страница 280: ...280 ...
Страница 294: ...294 Index ...
Страница 296: ...294 Chapter ...
Страница 297: ...H Manual Part Number G1099 90027 Copyright 1998 Hewlett Packard Printed in USA 11 98 ...