Maintaining the MSD
To lubricate the side plate O-ring
To lubricate the side plate O-ring
Materials needed:
Cloths, clean (05980-60051)
Gloves, clean, lint-free
large (8650-0030)
small (8650-0029)
Grease, Apiezon L, high vacuum (6040-0289)
The side plate O-ring needs a thin coat of grease to ensure a good vacuum seal. If
the side plate O-ring appears dry, or does not seal correctly, lubricate it using this
procedure.
1
Vent the MSD (page 52).
2
Open the vacuum manifold (page 54).
3
Use a clean, lint-free cloth or glove to spread a
thin
coat of high vacuum
grease on the exposed surface of the O-ring.
4
Use a clean, lint-free cloth or glove to wipe away excess grease.
If the O-ring looks shiny, there is too much grease on it.
5
Close the vacuum manifold (page 56).
6
Pump down the MSD (page 56).
& $ 8 7 , 2 1
Vacuum seals other than the side plate O-ring and vent valve O-ring do not need to be
lubricated. Lubricating other seals can interfere with their correct function.
& $ 8 7 , 2 1
Excess grease can trap air and dirt. Grease on surfaces of the O-ring other than the exposed
surface can trap air, resulting in air spikes during operation.
Содержание 5973
Страница 1: ...H Hardware Manual HP 5973 Mass Selective Detector ...
Страница 3: ...HP 5973 Mass Selective Detector Hardware Manual ...
Страница 25: ...1 Installing GC Columns To condition a capillary column 25 ...
Страница 30: ...30 ...
Страница 37: ...2 Operating the MSD To view MSD analyzer temperature and vacuum status 37 ...
Страница 39: ...2 Operating the MSD To set monitors for MSD temperature and vacuum status 39 ...
Страница 41: ...2 Operating the MSD To set the MSD analyzer temperatures 41 ...
Страница 43: ...2 Operating the MSD To set the interface temperature from the ChemStation 43 ...
Страница 51: ...2 Operating the MSD To remove the MSD covers 51 Upper cover Lower cover ...
Страница 92: ...92 ...
Страница 103: ...4 Maintaining the MSD To check and add foreline pump oil 103 Lower line Oil level window Fill cap Upper line ...
Страница 105: ...4 Maintaining the MSD To drain the foreline pump 105 Fill cap Drain plug Pump motor ...
Страница 107: ...4 Maintaining the MSD To refill the foreline pump 107 Fill cap Upper line Oil level window Drain plug ...
Страница 109: ...4 Maintaining the MSD To replace the oil trap 109 Oil trap Clamp O ring assembly Oil trap adapter Foreline pump ...
Страница 111: ...4 Maintaining the MSD To check the diffusion pump fluid 111 Fan Fluid level window sight glass ...
Страница 113: ...4 Maintaining the MSD To separate the MSD from the GC 113 ...
Страница 121: ...4 Maintaining the MSD To reconnect the MSD to the GC 121 ...
Страница 123: ...4 Maintaining the MSD To remove the calibration vial 123 Calibration valve Collar Calibration vial ...
Страница 125: ...4 Maintaining the MSD To refill and reinstall the calibration vial 125 Calibration valve Collar Calibration vial 6 mm ...
Страница 141: ...4 Maintaining the MSD To lubricate the side plate O ring 141 Side plate O ring ...
Страница 159: ...4 Maintaining the MSD To remove a filament 159 Filament 2 Filament 1 ...
Страница 161: ...4 Maintaining the MSD To reinstall a filament 161 Filament 2 Filament 1 ...
Страница 210: ...210 ...
Страница 211: ...6 GC MSD Interface This chapter describes the function of the GC MSD interface GC MSD interface 212 ...
Страница 214: ...214 ...
Страница 232: ...232 7 Analyzer Analyzer heaters and radiators ...
Страница 280: ...280 ...
Страница 294: ...294 Index ...
Страница 296: ...294 Chapter ...
Страница 297: ...H Manual Part Number G1099 90027 Copyright 1998 Hewlett Packard Printed in USA 11 98 ...