Detector
The detector in the MSD analyzer is a high energy conversion dynode (HED) cou-
pled to an electron multiplier (EM). The detector is located at the exit end of the
quadrupole mass filter. It receives the ions that have passed through the mass fil-
ter. The detector generates an electronic signal proportional to the number of ions
striking it. The detector has three main components: the detector focus lens, the
high energy dynode, and the electron multiplier horn.
Detector focus lens
The detector focus lens directs the ion beam into the HED, which is located off
axis. The voltage on the detector focus lens is fixed at -354 V.
High energy dynode
The high energy dynode (HED) operates at -10,000 volts, attracting the positive
sample ions exiting the quadrupole. The HED is located off-axis from the center of
the quadrupole mass filter to minimize signals due to photons, hot neutrals, and
electrons coming from the ion source. When the ion beam hits the HED, electrons
are emitted. These electrons are attracted to the more positive electron multiplier
horn.
Electron multiplier horn
The electron multiplier horn carries a voltage of up to -3000 volts at its opening
and 0 volts at the other end. The electrons emitted by the HED strike the EM horn
and cascade through the horn, liberating more electrons as they go. At the far end
of the horn, the current generated by the electrons is carried through a shielded
cable outside the analyzer to the signal amplifier board.
The voltage applied to the electron multiplier horn determines the gain. The volt-
age is adjustable from 0 to -3000 V dc. Use the electron multiplier voltage found
in autotune as a baseline for the electron multiplier voltage setting.
•
To increase signal strength, increase the electron multiplier voltage.
•
For concentrated samples where less signal strength is needed, decrease the
electron multiplier voltage.
Содержание 5973
Страница 1: ...H Hardware Manual HP 5973 Mass Selective Detector ...
Страница 3: ...HP 5973 Mass Selective Detector Hardware Manual ...
Страница 25: ...1 Installing GC Columns To condition a capillary column 25 ...
Страница 30: ...30 ...
Страница 37: ...2 Operating the MSD To view MSD analyzer temperature and vacuum status 37 ...
Страница 39: ...2 Operating the MSD To set monitors for MSD temperature and vacuum status 39 ...
Страница 41: ...2 Operating the MSD To set the MSD analyzer temperatures 41 ...
Страница 43: ...2 Operating the MSD To set the interface temperature from the ChemStation 43 ...
Страница 51: ...2 Operating the MSD To remove the MSD covers 51 Upper cover Lower cover ...
Страница 92: ...92 ...
Страница 103: ...4 Maintaining the MSD To check and add foreline pump oil 103 Lower line Oil level window Fill cap Upper line ...
Страница 105: ...4 Maintaining the MSD To drain the foreline pump 105 Fill cap Drain plug Pump motor ...
Страница 107: ...4 Maintaining the MSD To refill the foreline pump 107 Fill cap Upper line Oil level window Drain plug ...
Страница 109: ...4 Maintaining the MSD To replace the oil trap 109 Oil trap Clamp O ring assembly Oil trap adapter Foreline pump ...
Страница 111: ...4 Maintaining the MSD To check the diffusion pump fluid 111 Fan Fluid level window sight glass ...
Страница 113: ...4 Maintaining the MSD To separate the MSD from the GC 113 ...
Страница 121: ...4 Maintaining the MSD To reconnect the MSD to the GC 121 ...
Страница 123: ...4 Maintaining the MSD To remove the calibration vial 123 Calibration valve Collar Calibration vial ...
Страница 125: ...4 Maintaining the MSD To refill and reinstall the calibration vial 125 Calibration valve Collar Calibration vial 6 mm ...
Страница 141: ...4 Maintaining the MSD To lubricate the side plate O ring 141 Side plate O ring ...
Страница 159: ...4 Maintaining the MSD To remove a filament 159 Filament 2 Filament 1 ...
Страница 161: ...4 Maintaining the MSD To reinstall a filament 161 Filament 2 Filament 1 ...
Страница 210: ...210 ...
Страница 211: ...6 GC MSD Interface This chapter describes the function of the GC MSD interface GC MSD interface 212 ...
Страница 214: ...214 ...
Страница 232: ...232 7 Analyzer Analyzer heaters and radiators ...
Страница 280: ...280 ...
Страница 294: ...294 Index ...
Страница 296: ...294 Chapter ...
Страница 297: ...H Manual Part Number G1099 90027 Copyright 1998 Hewlett Packard Printed in USA 11 98 ...