4
Maintaining the MSD
99
Chemical residue is another potential danger
Only a small portion of your sample is ionized by the ion source. The
majority of any sample passes through the ion source without being ionized.
It is pumped away by the vacuum system. As a result, the exhaust from the
foreline pump will contain traces of the carrier gas and your samples. It will
also contain tiny droplets of foreline pump oil.
An oil trap is supplied with the foreline pump. This trap stops
only
pump
oil droplets. It
does not
trap any other chemicals. If you are using toxic
solvents or analyzing toxic chemicals, do not use the oil trap. Instead, install
a hose to take the exhaust from the foreline pump outdoors or into a fume
hood vented to the outdoors. Be sure to comply with your local air quality
regulations.
The fluids in the diffusion pump and foreline pump also collect traces of the
samples being analyzed. All used pump fluid should be considered hazard-
ous and handled accordingly. Dispose of used fluid correctly, as specified
by your local regulations.
: $ 5 1 , 1 *
7KHRLOWUDSVWRSVRQO\IRUHOLQHSXPSRLO,WGRHVQRWWUDSRUILOWHURXWWR[LFFKHPLFDOV,I\RX
DUHXVLQJWR[LFVROYHQWVRUDQDO\]LQJWR[LFFKHPLFDOVUHPRYHWKHRLOWUDS'RQRWXVHWKHWUDS
LI\RXKDYHD&,06',QVWDOODKRVHWRWDNHWKHIRUHOLQHSXPSH[KDXVWRXWVLGHRUWRDIXPH
KRRG
: $ 5 1 , 1 *
:KHQUHSODFLQJSXPSIOXLGXVHDSSURSULDWHFKHPLFDOUHVLVWDQWJORYHVDQGVDIHW\JODVVHV
$YRLGDOOFRQWDFWZLWKWKHIOXLG
Содержание 5973
Страница 1: ...H Hardware Manual HP 5973 Mass Selective Detector ...
Страница 3: ...HP 5973 Mass Selective Detector Hardware Manual ...
Страница 25: ...1 Installing GC Columns To condition a capillary column 25 ...
Страница 30: ...30 ...
Страница 37: ...2 Operating the MSD To view MSD analyzer temperature and vacuum status 37 ...
Страница 39: ...2 Operating the MSD To set monitors for MSD temperature and vacuum status 39 ...
Страница 41: ...2 Operating the MSD To set the MSD analyzer temperatures 41 ...
Страница 43: ...2 Operating the MSD To set the interface temperature from the ChemStation 43 ...
Страница 51: ...2 Operating the MSD To remove the MSD covers 51 Upper cover Lower cover ...
Страница 92: ...92 ...
Страница 103: ...4 Maintaining the MSD To check and add foreline pump oil 103 Lower line Oil level window Fill cap Upper line ...
Страница 105: ...4 Maintaining the MSD To drain the foreline pump 105 Fill cap Drain plug Pump motor ...
Страница 107: ...4 Maintaining the MSD To refill the foreline pump 107 Fill cap Upper line Oil level window Drain plug ...
Страница 109: ...4 Maintaining the MSD To replace the oil trap 109 Oil trap Clamp O ring assembly Oil trap adapter Foreline pump ...
Страница 111: ...4 Maintaining the MSD To check the diffusion pump fluid 111 Fan Fluid level window sight glass ...
Страница 113: ...4 Maintaining the MSD To separate the MSD from the GC 113 ...
Страница 121: ...4 Maintaining the MSD To reconnect the MSD to the GC 121 ...
Страница 123: ...4 Maintaining the MSD To remove the calibration vial 123 Calibration valve Collar Calibration vial ...
Страница 125: ...4 Maintaining the MSD To refill and reinstall the calibration vial 125 Calibration valve Collar Calibration vial 6 mm ...
Страница 141: ...4 Maintaining the MSD To lubricate the side plate O ring 141 Side plate O ring ...
Страница 159: ...4 Maintaining the MSD To remove a filament 159 Filament 2 Filament 1 ...
Страница 161: ...4 Maintaining the MSD To reinstall a filament 161 Filament 2 Filament 1 ...
Страница 210: ...210 ...
Страница 211: ...6 GC MSD Interface This chapter describes the function of the GC MSD interface GC MSD interface 212 ...
Страница 214: ...214 ...
Страница 232: ...232 7 Analyzer Analyzer heaters and radiators ...
Страница 280: ...280 ...
Страница 294: ...294 Index ...
Страница 296: ...294 Chapter ...
Страница 297: ...H Manual Part Number G1099 90027 Copyright 1998 Hewlett Packard Printed in USA 11 98 ...