246
8
Electronics
Back panel and connectors
High vacuum signal (
HIVAC SIGNAL
) connector
The high vacuum signal connector is on the ac board, extending to the back panel.
The high vacuum signal cable carries control signals to the turbo controller from
the ac board, and high vacuum pump status information to the ac board. This
cable also carries control and power to the calibration valve and the high-vacuum
cooling fan.
High vacuum power (
HIVAC POWER
) connector
The high vacuum power connector carries power for the diffusion pump heater or
the turbo controller. The high vacuum power jumper connects the ac board to the
high vacuum power connector.
Power cord receptacle
The ac power cord brings in all electrical power for the MSD. The power cord can
be detached from the MSD. The ac power receptacle is on the ac board, extending
through the back panel.
Foreline pump power cord receptacle
The foreline pump power cord receptacle provides ac power for the foreline pump.
If the power switch is off, no power is supplied to the foreline pump. The ac power
receptacle is on the ac board, extending through the back panel.
Primary fuses
The primary fuses limit current into the MSD in case of a short circuit in the fore-
line pump. The primary fuses are on the ac board, extending through the back
panel.
Содержание 5973
Страница 1: ...H Hardware Manual HP 5973 Mass Selective Detector ...
Страница 3: ...HP 5973 Mass Selective Detector Hardware Manual ...
Страница 25: ...1 Installing GC Columns To condition a capillary column 25 ...
Страница 30: ...30 ...
Страница 37: ...2 Operating the MSD To view MSD analyzer temperature and vacuum status 37 ...
Страница 39: ...2 Operating the MSD To set monitors for MSD temperature and vacuum status 39 ...
Страница 41: ...2 Operating the MSD To set the MSD analyzer temperatures 41 ...
Страница 43: ...2 Operating the MSD To set the interface temperature from the ChemStation 43 ...
Страница 51: ...2 Operating the MSD To remove the MSD covers 51 Upper cover Lower cover ...
Страница 92: ...92 ...
Страница 103: ...4 Maintaining the MSD To check and add foreline pump oil 103 Lower line Oil level window Fill cap Upper line ...
Страница 105: ...4 Maintaining the MSD To drain the foreline pump 105 Fill cap Drain plug Pump motor ...
Страница 107: ...4 Maintaining the MSD To refill the foreline pump 107 Fill cap Upper line Oil level window Drain plug ...
Страница 109: ...4 Maintaining the MSD To replace the oil trap 109 Oil trap Clamp O ring assembly Oil trap adapter Foreline pump ...
Страница 111: ...4 Maintaining the MSD To check the diffusion pump fluid 111 Fan Fluid level window sight glass ...
Страница 113: ...4 Maintaining the MSD To separate the MSD from the GC 113 ...
Страница 121: ...4 Maintaining the MSD To reconnect the MSD to the GC 121 ...
Страница 123: ...4 Maintaining the MSD To remove the calibration vial 123 Calibration valve Collar Calibration vial ...
Страница 125: ...4 Maintaining the MSD To refill and reinstall the calibration vial 125 Calibration valve Collar Calibration vial 6 mm ...
Страница 141: ...4 Maintaining the MSD To lubricate the side plate O ring 141 Side plate O ring ...
Страница 159: ...4 Maintaining the MSD To remove a filament 159 Filament 2 Filament 1 ...
Страница 161: ...4 Maintaining the MSD To reinstall a filament 161 Filament 2 Filament 1 ...
Страница 210: ...210 ...
Страница 211: ...6 GC MSD Interface This chapter describes the function of the GC MSD interface GC MSD interface 212 ...
Страница 214: ...214 ...
Страница 232: ...232 7 Analyzer Analyzer heaters and radiators ...
Страница 280: ...280 ...
Страница 294: ...294 Index ...
Страница 296: ...294 Chapter ...
Страница 297: ...H Manual Part Number G1099 90027 Copyright 1998 Hewlett Packard Printed in USA 11 98 ...