Chapter 3 Basic Operation
3.4 Measurement
F-74BW
73
3.4.3 Auto Data Memory
The instrument can perform data memory automatically
at a regular interval. Refer to "3.2.3 Auto Data Memory
Setting" P.20 for the setting operations.
If the auto power off setting is ON, it is temporarily OFF
while this function is active.
1.
After the interval setting is completed, press the
MEAS key to return to the measurement screen.
"Auto LOG" is displayed on the screen.
2.
Press the ENTER key.
"Auto LOGGING" is displayed on the screen, and
the auto data memory function starts.
3.
To stop the auto data memory, press the ENTER
key.
・
Do not power ON/OFF while the auto data memory function
works.
It may damage the reliability of the memory data.
・
Only the POWER, ENTER, and the
▲
or
▼
keys are possible
while the auto data memory function works.
・
The actual date of memory may have the difference of ±1
second from the time calculated based on the interval setting.
・
When the number of memory data exceeds 999, data memory
is stopped and ERROR No.0010 is displayed.