Chapter 4 Maintenance and Troubleshooting
4.3 Conductivity Electrode Maintenance
F-74BW
99
4.3
Conductivity Electrode Maintenance
Refer to the electrode operation manuals for the mainte-
nance operations of each electrode.
Long-term storage
When an electrode will not be used for a long period of
time, store it after performing the following procedures.
Also, perform the maintenance of the electrode every 3
to 6 months.
1.
Disconnect the electrode from the instrument.
2.
Use pure water to wash away any sample
solution that may be adhered to the electrode.
3.
Wash the inside of the electrode protective cap
with pure water, then, after shaking out the
water, fill the cap with enough pure water to
soak the sponge.
4.
Place the electrode protective cap on the
electrode.