Chapter 4 Maintenance and Troubleshooting
4.2 ION Electrode Maintenance
F-74BW
97
Filling the electrode: assembly
1.
Return the silicon tube to the original position to
seal the internal solution filler port (inner tube).
Make sure that the filler port is completely sealed.
2.
If the bodies 1 or 2 or the liquid junction are dirty,
wash them with pure water (deionized).
3.
Insert the body 1 into the body 2.
Make sure that the O-ring is securely sealed.
4.
Return the spring to the top of body 1.
5.
Hold the rubber sleeve in place by hand, then twist
the electrode protective cap at 90
°
and fit the rubber
sleeve into the electrode protective cap.
6.
Adjust the HORIBA logo on the electrode protective
cap and the filling port of the outer tube face in the
same direction and fit the cap into the body 2.
7.
Twist the rubber sleeve at 90
°
and hold it in place.
8.
Use a dropper to fill the electrode with the specified
internal solution.
9.
Make the internal solution seep out from the liquid
junction referring to "
●
Daily maintenance" P.94.
10.
Store the electrode referring to "
●
Short-term
storage" P.93.