Chapter 4 Maintenance and Troubleshooting
4.4 Troubleshooting
110
HORIBA
●
The measured value is blinks.
・
The pH value is outside the measurement range
(when pH value is displayed).
Measurement range: pH0.000 to pH14.000
・
The mV value is outside the display range
(when mV value is displayed).
Display range: ±1999.9 mV
・
The ion value is outside the display range
(when ion value is displayed).
Display range: 99.9 or more
・
The conductivity value is outside the display range
(when conductivity value is displayed).
Display range:
0.00 to 19.99 (when cell constant is 100 m
-1
)
Cause
Countermeasure
The sample is
inappropriate.
Use a sample within the measurement
range.
The liquid junction
is not immersed in
the sample.
Immerse the electrode up to 3 cm from its
tip.
The electrode
cable is broken.
Replace the electrode.
The instrument is
defective.
Perform the check operation mentioned
below.
Calibration is
unperformed or
incorrect.
Perform calibration correctly.