Chapter 3 Basic Operation
3.2 Settings
F-74BW
21
3.2.4 pH Calibration Interval Setting
This setting allows you to ensure calibration at a regular
interval. When the set inteval (days) elapses after the
previous calibration, ERROR No.0008 (Calibration cycle
error) is displayed.
This setting is possible only for CH1.
1.
Press the SET key on the measurement screen to
enter the pH calibration interval setting screen.
2.
Press the ENTER key to enter the calibration inter-
val change screen.
3.
Press the
or
key to change ON or OFF, and
press the ENTER key.
4.
If you select ON, press the
or
key to select a
digit and press the
▲
or
▼
key to change the value,
and then press the ENTER key.
Setting range: 1 to 999 days
▲
▼
▲
▼