Chapter 3 Basic Operation
3.2 Settings
24
HORIBA
3.2.7 ION Valency Setting
Select the valency of the ion to be measured.
If you use different ion valency settings between
calibration and measurement, you can not obtain correct
measurement results. When you change the ion valency
setting, perform calibration again.
This setting is possible only for CH1.
1.
Press the SET key on the measurement screen,
and press the
▲
or
▼
key to enter the ion valency
setting screen.
2.
Press the ENTER key to enter the ion valency
change screen.
3.
Select the ion valency, and press the ENTER key.