Chapter 3 Basic Operation
3.3 Calibration
F-74BW
67
7.
Check the stable indication value, and press the
CAL key.
The HOLD indicator blinks until the reading
stabilizes. When the reading stabilizes, the HOLD
indicator lights up and the reading is held.
8.
After the calibration is completed, press the MEAS
key to return to the measurement screen.
3.3.4 Salinity Calibration
You can perform 1 point calibration using a salinity stan-
dard solution.
1.
Prepare the electrode.
Wash the electrode with pure water (deionized),
and wipe it off by filter paper or tissue paper.
2.
Immerse the electrode in the desired standard
solution used for calibration.
3.
Press the
▲
or
▼
key when the measurement
screen is displayed to light up ACT for CH2.
4.
Press the MODE key on the measurement screen
to enter the salt measurement screen.
5.
Press the CAL key to enter the salinity calibration
screen.