Chapter 4 Maintenance and Troubleshooting
4.4 Troubleshooting
106
HORIBA
●
ERROR No.0011 Cell constant out of range
Definition
The cell constant is out of the automatic calculation
range.
●
ERROR No.000E Key operation ERROR
Definition
Key operation is invalid.
Cause
Countermeasure
COND electrode is at
the end of the useful
life.
Replace the electrode.
Improper standard
solution
Prepare new standard solution.
Cause
Countermeasure
Loose connection or
breaking of keys
Disconnect and reconnect the AC
adapter, and press the ON/OFF key 10
seconds later.
If the same error occurs again after the
countermeasure is performed, contact
HORIBA sales representative.