![horiba F-74BW Скачать руководство пользователя страница 104](http://html1.mh-extra.com/html/horiba/f-74bw/f-74bw_instruction-manual_2162212104.webp)
Chapter 4 Maintenance and Troubleshooting
4.2 ION Electrode Maintenance
92
HORIBA
4.2
ION Electrode Maintenance
Refer to the electrode operation manuals for the mainte-
nance operations of each electrode.
●
Ion electrodes
(*):
Change in the electric potential of the electrode (25ºC)
when the ion concentration is changed by a factor of 10.
The above electrodes are subject to change without notice.
The maintenance operations of the ion electrodes listed
in the above table are describes below.
Ion to be
measured
Ion type
Slope(*)
Measurement
range
Elect-
rode
model
Compati
ble tip
model
Internal solution
Potassium
K
+
+1
+58 mV
0.04 to 39,000
mg/L
6582
-10C
7682
3.33 mol/L
NaCl
Calcium
Ca
2+
+2
+29 mV
0.40 to 40,080
mg/L
6583
-10C
7683
3.33 mol/L
KCl (model 300)
Chloride
CI
-
−
1
−
59 mV
0.4 to 35,000
mg/L
6560
-10C
7660
1 mol/L
KNO
3
Fluoride
F
-
−
1
−
59 mV
0.02 to 19,000
mg/L
6561
-10C
7661
3.33 mol/L
KCl (model 300)
Nitrate
NO
3
-
−
1
−
55 mV
0.62 to 62,000
mg/L
6581
-10C
7681
1 mol/L
KCl
Ammonia
NH
3
+1
−
59 mV
0.1 to 1,000
mg/L
5002A
-10C
Included inter-
nal solution
NH
4
Cl