Chapter 4 Maintenance and Troubleshooting
4.4 Troubleshooting
F-74BW
111
Check operation
●
The temperature display blinks.
The temperature display is fixed at 25
°
C.
The temperature measurement value is outside the
measurement range.
Measurement range:
−
30
°
C to 130
°
C
①
pH, mV, ion, resistivity
Short the connecter for the electrode on the
instrument between the center pin and outer
metal.
If the measurement value stops blinking in this
state, the instrument is normal.
②
Conductivity, salinity, TDS
Disconnect the electrode.
If the displayed measurement value is near 0, the instrument is
normal.
Cause
Countermeasure
The temperature of
the sample is out-
side display range.
Check the sample temperature.
Adjust it within the measurement range.
The internal ther-
mistor connection
of the electrode is
broken or shorted.
Check the resistance of the temperature
sensor connector. If it is 50 k
Ω
or more at
room temperature, replace the electrode.
The connection of
the temperature
connector is failed.
Insert the temperature connector firmly.
The instrument is
defective.
Contact HORIBA sales representative.
The settings of the
temperature cali-
bration is failed.
Initialize the settings referring to "
●
Setting Initialization" P.42.