![horiba F-74BW Скачать руководство пользователя страница 61](http://html1.mh-extra.com/html/horiba/f-74bw/f-74bw_instruction-manual_2162212061.webp)
Chapter 3 Basic Operation
3.3 Calibration
F-74BW
49
2.
Open the filler port of the internal solution. During
the calibration, the filler port should be kept open.
3.
Immerse the pH electrode in the standard solution
of pH7 to calibrate the first point.
4.
Press the
▲
or
▼
key on the measurement
screen is displayed to
light up ACT for CH1.
5.
Press the CAL key to enter the USA calibration
screen.
6.
Check the stable indication value, and press the
CAL key.
The HOLD indicator blinks until the reading
stabilizes. When the reading stabilizes, the HOLD
indicator lights up and the reading is held.
The
mark, which means that the pH7
calibration is completed, is displayed on the screen.
・
To stop calibration:
Press the CAL key while the HOLD indicator
blinks.
Содержание F-74BW
Страница 1: ...Instruction Manual pH ION COND METER F 74BW...
Страница 11: ...CONTENTS IX Chapter 5 Specifications 113 5 1 Specifications 113 5 2 Default Settings 116 5 3 Options 117...
Страница 12: ......
Страница 130: ......
Страница 132: ......