Chapter 4 Maintenance and Troubleshooting
4.4 Troubleshooting
108
HORIBA
When there is a problem in the sample
●
The response is slow.
Cause
Countermeasure
The liquid junction
is not immersed in
the sample solu-
tion.
Immerse the electrode in the sample solu-
tion up until the liquid junction.
Caused by the
effects of the sam-
ple.
Judge by measuring with a stable stan-
dard solution.
Contact HORIBA sales representative.
Cause
Countermeasure
Caused by the
effects of the sam-
ple.
Response time may slow down, depend-
ing on the properties of the sample solu-
tion.
Contact HORIBA sales representative.
Wash the respon-
sive membrane.
Wash the responsive membrane.
The responsive
membrane is dam-
aged or worn out.
Replace the electrode.
There is a problem
with the standard
solution.
Fill the electrode with new standard solu-
tion, as noted in the electrode
operation manual.