CM4 Four-Point Continuous Monitor
CM4 Technical Manual
7-11
recent concentration reading. However, it will not report any changes in
the concentration during this period, because the tape is saturated.
The Monitor relay (RY6) will energize when monitoring is temporarily
suspended due to tape saturation. Note however, any one point
that is enabled for Duty Cycle will cause the Monitor relay (RY6) to
energize regardless if some points have not been selected (enabled)
for Duty Cycle. In the CM4 software, the “monitor relay action during
Duty Cycle” bits default to TRUE, but can be set to FALSE with the
following procedure.
Instruction
If the Duty Cycle option is installed, you can use Menu Function 3.3.6
(EN/DIS Duty Cycle) to enable or disable the Duty Cycle function for
each point. Use Menu Function 3.5.6 (Set Duty Cycle) to program the
Duty Cycle time function.
Follow these steps to enable/disable the Duty Cycle for each point:
1. Press PROGRAM to exit Monitor Mode.
2. From the 3–SET-UP MENU, choose the subfunction 3–OUTPUT
OPTIONS, then 6–EN/DIS DUTY CYCLE.
3 – SET-UP MODE
3 – OUTPUT OPTION
6 – EN/DIS DUTY CYCLES
NOTE:
Duty Cycle is a factory-enabled option. If not set at the factory,
then it is not available to user.
3. The display reads:
SELECT POINT
Use the numeric keys to specify a point.
4. The display will read:
PT 1 ENABLE 1=Y 0=N
Choose 1 or 0 to enable or disable that point for Duty Cycle.
5. S e l e c t o t h e r p o i n t s , i f d e s i r e d , o r p r e s s
RESET/ESCAPE key.
7.5 Function 3.4 – Configuration
Purpose
Use the configuration function to:
• Save new instrument configurations
• Restore an existing instrument configuration after changing
configuration parameters or a RESET SYSTEM that returned
parameters to the factory default
System configuration consists of all configurable parameters defined
in the Set-Up Menu.
Instructions
1. Press PROGRAM to exit Monitor Mode.
2. From the 3–SET-UP MENU, choose the subfunction
4–CONFIGURATION.
3 – SET-UP MODE
4 – CONFIGURATION
The display reads:
1=SAVE 2=LOAD CONFIG