CM4 Four-Point Continuous Monitor
CM4 Technical Manual
B-7
Fault
No.
Description
Possible Cause
Corrective Action
Relay
Activated
Beeping
Audio
Alarm
Yellow
Status
LEDs
43
Chemcassette
®
Stale
Chemcassette
®
past “out-of-bag” expiration
date (continues monitoring - no accuracy
guarantee) (See
)
Replace Chemcassette
®
MAINT
(RY8)
No
44
Wrong Chemcassette
®
Type
Wrong Chemcassette
®
type for gas family Replace with correct Chemcassette
®
FAULT
(RY10)
On
Solid
45
Chemcassette
®
Past Expiry
Chemcassette
®
past expiration date (stops
monitoring) (See
)
Replace Chemcassette
®
FAULT
(RY10)
On
Solid
46
Chemcassette
®
Very Stale
Chemcassette
®
past “out-of-bag” expiration
date (stops monitoring) (See
Replace Chemcassette
®
FAULT
(RY10)
On
Solid
Table B-1: CM4 Fault Listings (cont’d)
珠海司福斯特科技有限公司
www.safedtech.com