CM4 Four-Point Continuous Monitor
CM4 Technical Manual
9-5
CM4 OPTICS TEST CARD
INSTRUCTIONS
FOR OPTICS VERIFICATION PROCEDURE,
ENTER CALIBRATION MODE 1.1, OPTICS VERIFY
SEE MANUAL FOR DETAILS.
STAIN 1
INSER
T
Figure 9-2: Optics Verification Card
9.2.3 Loading the Carrier
To load the Chemcassette
®
carrier:
1. Open the carrier.
2. Remove the supply reel and discard any remaining
Chemcassette
®
tape. Remove and discard the entire take-up
reel.
3. Using your thumb, press the center of the supply reel until the
top half separates from the bottom half.
4. Move the bottom (slotted side) of the reel to the take-up
position.
5. Place the new supply reel into position and thread the
Chemcassette
®
tape through the slot in the carrier and into
the slot on the take-up reel.
6. Turn the take-up reel 3 to 4 times to advance the Chemcassette
®
tape (2) and take up slack.
7. Reinstall the top half of the take-up reel until a click is heard.
8. Close the carrier.
3
2
5
5
4
1
Figure 9-3: Loading the Carrier
Legend
1
Carrier
4
Take-up reel
2
Chemcassette
®
Tape
5
Tape guide
3
Supply reel