219
Error Display
Display
Meaning
Solution
• Error occurred during file
processing.
Either use another USB memory, or
back up the files on USB memory
and then format and use.
• Measurement data was saved
without measuring.
• Measuring started while
the measurement data was
being saved to file, so the
measurement data that was
being saved was lost.
Make the measurements again, and
then save the measurement data.
Other errors
Display
Meaning
Solution
• The internal temperature
exceeded the operating range,
or the cooling fan stopped.
Turn OFF the power supply
immediately, and check the
instrument installation environment
and cooling fan status.
If the situation still does not improve,
there is a risk of malfunction. Contact
your authorized Hioki distributor or
reseller.
Displayed in the following cases.
• If the power supply switch was
turned OFF while the settings
were being saved (including
power being lost due to a power
failure, etc.)
• If the settings were initialized
due to a version upgrade (When
the power supply is turned ON
after a version upgrade)
• Reconfigure this setting after
initializing the instrument’s settings.
• If an error is displayed even after
the settings have been redone,
there is a risk of malfunction.
Contact your authorized Hioki
distributor or reseller.
If you encounter an error that is not listed above, please make a note of the display and contact
your authorized Hioki distributor or reseller.
14
Maintenance and Service
Содержание ST4030
Страница 2: ......
Страница 14: ...6 Usage Notes ...
Страница 32: ...24 Inspecting before Measurement and Verifying Operations ...
Страница 42: ...34 Table Copying ...
Страница 108: ...100 Setting Dielectric Break down Judgment Conditions ...
Страница 132: ...124 Terminal Open Error Setting ...
Страница 198: ...190 Editing Files and Folders ...
Страница 214: ...206 Interface Specifications ...
Страница 237: ...229 Dimensions Diagram 15 4 Dimensions Diagram Unit mm 215 1 348 3 178 0 5 285 6 0 5 200 1 13 40 3 1 15 3 1 15 Appendix ...
Страница 248: ...240 Precautions when Processing Test Leads ...
Страница 252: ...244 Index ...
Страница 253: ......
Страница 254: ......
Страница 255: ......
Страница 256: ......