194
Functional Specifications
(1) LC and RC value judgments
LC and RC values judgments
[LCRC AREA]
Compares the LC and RC values distribution of the master workpiece and the
LC and RC values of the workpiece to be tested.
LC and RC values: The quantification of the response waveform as 2
parameters of the LC and RC values
1. Sets the waveform area used in LC and RC values operations.
Automatically set from the master waveform during voltage calibration.
Settable using communications commands if you want to set your own
operations area.
Settings range: Within sampling data range
2. Imports standard LC and RC values to create the PASS judgment area
after implementing voltage calibration.
3. Sets the PASS judgment area using the imported standard LC and RC
values.
Automatic creation: Automatically creates the PASS judgment area using
the imported standard LC and RC values.
Manual creation: Manually sets the PASS judgment area
Sets the peaks of the PASS judgment area (square) using the LC and RC
values.
Settings range: ±(0.000 f to 1.000) (valid to 4 digits)
(2) Waveform judgments
Waveform surface area
comparison judgment
[AREA]
Compares and judges the master waveform surface area and the test waveform
surface area in the area specified by the user.
Judgment values
Sets the surface area ratio threshold value
Settings range: OFF / 0.00 to 99.99% (0.01% steps)
Judgment area
Sets the judgment area using points
Settings area: Within sampling data range
Waveform difference surface
area comparison judgments
[DIFF-AREA]
Calculates and judges the ratios of the master waveform and test waveform
differences over the surface area of the master waveform
Judgment values
Sets the difference threshold value
Settings range: OFF / 0.00 to 99.99% (0.01% steps)
Judgment area
Sets the judgment area using points
Settings area: Within sampling data range
Waveform flutter detection
judgments
[FLUTTER]
The amount of the high-frequency component expressed by the test waveform
in the user-designated area is detected, and judge the discharge by its size.
Primarily differentiates the test waveform, and calculates the surface area of the
differentiated waveform as the discharge component amount.
Flutter
=
Number of data
Σ
n
=1
|
d
n
−
d
n
−1
|
Judgment values
Sets the high-frequency component amount
Settings range: OFF / 0 to 999,999 (1 step)
Judgment area
Sets the judgment area using points
Settings area: Within sampling data range
Содержание ST4030
Страница 2: ......
Страница 14: ...6 Usage Notes ...
Страница 32: ...24 Inspecting before Measurement and Verifying Operations ...
Страница 42: ...34 Table Copying ...
Страница 108: ...100 Setting Dielectric Break down Judgment Conditions ...
Страница 132: ...124 Terminal Open Error Setting ...
Страница 198: ...190 Editing Files and Folders ...
Страница 214: ...206 Interface Specifications ...
Страница 237: ...229 Dimensions Diagram 15 4 Dimensions Diagram Unit mm 215 1 348 3 178 0 5 285 6 0 5 200 1 13 40 3 1 15 3 1 15 Appendix ...
Страница 248: ...240 Precautions when Processing Test Leads ...
Страница 252: ...244 Index ...
Страница 253: ......
Страница 254: ......
Страница 255: ......
Страница 256: ......