Troubleshooting
Symptom
Check item or cause
Corrective action and reference page
Error occurs during testing
Have you checked the details
of the error icon displayed
onscreen?
Check the details of the error icon
displayed onscreen.
Reference: “14.3 Error Display” (p. 214)
Is probing unstable?
An open error such as unstable probing
may occur. Stabilize the probing status.
Reference: “14.3 Error Display” (p. 214)
Is
[UNCAL]
displayed?
Calibrate the voltages.
Reference: “4.6 Voltage Calibration”
Error occurs during voltage
calibration
Have you checked the details
of the error window displayed
onscreen?
Check the details of the error window
displayed onscreen.
Reference: “14.3 Error Display” (p. 214)
Is the applied voltage too high?
If the applied voltage is too high, the
workpiece discharge will cause an error.
Lower the applied voltage to a suitable
voltage.
Reference: “14.3 Error Display” (p. 214)
Is the sampling frequency too
fast?
Unless a sufficient number of waveforms
is sampled during the test, an insufficient
number of waveforms will occur. Slow
the sampling frequency and make sure
that a sufficient number of waveforms is
sampled.
Reference: “14.3 Error Display” (p. 214)
Is the workpiece inductance
value too low?
If the set voltage is too high for the
workpiece inductance value, it may not
be possible to apply the set voltage
to the data. Set a suitable application
voltage for the workpiece inductance
value.
Reference: “15.6 Max. Applicable
Voltage (Reference Value) According to
the Workpiece Inductance” (p. 232)
LCD is blurred
Are you pressing the LCD too
forcefully?
Press the LCD more gently. Although
the blur may be considerable, there is
no functional problem.
Содержание ST4030
Страница 2: ......
Страница 14: ...6 Usage Notes ...
Страница 32: ...24 Inspecting before Measurement and Verifying Operations ...
Страница 42: ...34 Table Copying ...
Страница 108: ...100 Setting Dielectric Break down Judgment Conditions ...
Страница 132: ...124 Terminal Open Error Setting ...
Страница 198: ...190 Editing Files and Folders ...
Страница 214: ...206 Interface Specifications ...
Страница 237: ...229 Dimensions Diagram 15 4 Dimensions Diagram Unit mm 215 1 348 3 178 0 5 285 6 0 5 200 1 13 40 3 1 15 3 1 15 Appendix ...
Страница 248: ...240 Precautions when Processing Test Leads ...
Страница 252: ...244 Index ...
Страница 253: ......
Страница 254: ......
Страница 255: ......
Страница 256: ......