Error Display
Display *
1
EXT.I/O
ERR pin
Communications
VCAL status
Solution
ERROR 23
Anomalous value detected
ON
8
• Model ST4030 does not support the
workpiece.
• There is a risk that the probing status is
unstable.
• If you encounter this error with a normal,
non-defective workpiece, the instrument
may be malfunctioning. Please contact your
authorized Hioki distributor or reseller.
ERROR 24
Waveform analysis failure
ON
9
• Model ST4030 does not support the
workpiece.
• There is a risk that the probing status is
unstable.
• If you encounter this error with a normal,
non-defective workpiece, the instrument
may be malfunctioning. Please contact your
authorized Hioki distributor or reseller.
ERROR 25
Voltage set as the application voltage not
reached
ON
3
Cannot apply the set voltage to the workpiece.
If the set voltage is too high for the workpiece
inductance value, it may not be possible to
apply the set voltage. Set a suitable application
voltage for the workpiece inductance value.
Reference: “15.6 Max. Applicable Voltage
(Reference Value) According to the Workpiece
ERROR 27
Overrange value detected
ON
10
• There is a risk that the probing status is
unstable. Perform voltage calibration after
establishing stable probing.
• Check that the workpiece is normal.
The oscillation frequency of the detection
waveform is too long to reach the
waveform length
ON
4
• The length of the imported waveform is
insufficient. Set the sampling frequency and
number of sampling data so that at least 4
waves or more are imported.
Reference: “4.4 Sampling Frequency/Number
• Under some conditions, this error may occur
with Ver. 1.01-compatible tables.
Excessive waveform length relative to the
detected waveform’s oscillation period
ON
5
• The waveform has shrunk too much due to the
excessive length of the captured waveform.
Reduce the wave number by setting the
sampling frequency and sampling data count.
Reference: “4.4 Sampling Frequency/Number
• Under some conditions, this error may occur
with Ver. 1.01-compatible tables.
*1: If the voltage was calibrated from communications, the window is not displayed.
Содержание ST4030
Страница 2: ......
Страница 14: ...6 Usage Notes ...
Страница 32: ...24 Inspecting before Measurement and Verifying Operations ...
Страница 42: ...34 Table Copying ...
Страница 108: ...100 Setting Dielectric Break down Judgment Conditions ...
Страница 132: ...124 Terminal Open Error Setting ...
Страница 198: ...190 Editing Files and Folders ...
Страница 214: ...206 Interface Specifications ...
Страница 237: ...229 Dimensions Diagram 15 4 Dimensions Diagram Unit mm 215 1 348 3 178 0 5 285 6 0 5 200 1 13 40 3 1 15 3 1 15 Appendix ...
Страница 248: ...240 Precautions when Processing Test Leads ...
Страница 252: ...244 Index ...
Страница 253: ......
Страница 254: ......
Страница 255: ......
Страница 256: ......