214
Error Display
14.3
Error Display
If an error is displayed, repairs are required. Contact your authorized Hioki distributor or reseller.
Error during tests
Error type
EXT.I/O
ERR pin
Communications
measurement
status
Judgment
results
Total
judgment
results
Solution
Normal
OFF
0
Normal
judgment
Normal
judgment
–
Unmeasured /
Measurement stopped
OFF
1
No
judgments
No
judgments
–
Terminal open error
ON
2
OUT
FAIL
There is a risk of being unable to
contact the workpiece. Check the
probing status.
UNCAL error *
2
ON
4
OUT
FAIL
This happens when
“implementing a test with a table
that has not yet implemented
a voltage calibration” has been
selected. Calibrate the voltages
in test conditions settings mode.
Over-range error
ON
6
OUT
FAIL
A voltage that exceeded the input
voltage range was detected.
Either the workpiece’s electrical
characteristics differ markedly
from those of the master
workpiece, or the instrument
may not be in contact with the
workpiece. Check the workpiece
and probing state.
Voltage error
ON
6
OUT
FAIL
The magnitude of the voltage rise
was less than expected, with the
result that either the maximum
voltage was not reached, or the
instrument was unable to apply
the set voltage to the workpiece.
If the set voltage is too high
relative to the workpiece’s
inductance value, the instrument
may be unable to apply the
set voltage. Set an application
voltage that is suited to the
workpiece’s inductance value.
Overcurrent error
ON
7
OUT
FAIL
An excessive output current was
detected. This error may occur if
the workpiece is shorted.
Invalid table error
OFF *
1
1
No
judgments
No
judgments
The table’s settings have not
been initialized. Choose a table
with valid settings.
Содержание ST4030
Страница 2: ......
Страница 14: ...6 Usage Notes ...
Страница 32: ...24 Inspecting before Measurement and Verifying Operations ...
Страница 42: ...34 Table Copying ...
Страница 108: ...100 Setting Dielectric Break down Judgment Conditions ...
Страница 132: ...124 Terminal Open Error Setting ...
Страница 198: ...190 Editing Files and Folders ...
Страница 214: ...206 Interface Specifications ...
Страница 237: ...229 Dimensions Diagram 15 4 Dimensions Diagram Unit mm 215 1 348 3 178 0 5 285 6 0 5 200 1 13 40 3 1 15 3 1 15 Appendix ...
Страница 248: ...240 Precautions when Processing Test Leads ...
Страница 252: ...244 Index ...
Страница 253: ......
Страница 254: ......
Страница 255: ......
Страница 256: ......