213
Initializing the Instrument (Full Reset)
14.2 Initializing the Instrument (Full Reset)
Implementing a full reset will return all settings to their factory default state.
Implement a full reset only in the following circumstances.
• If the regular reset screen cannot be displayed due to a instrument error
Reference: “8.11 Initializing the Instrument (System Reset)” (p. 121)
Implement self-test after a full reset to check that there are no errors.
Reference: “9.2 Self-test Function” (p. 126)
• If you have forgotten the interlock or key lock passwords
IMPORTANT
• Remove the test leads from the workpiece and implement a full reset.
• If operations are still not normal even after a full reset, repairs are required. Contact your authorized Hioki
distributor or reseller.
Lights red
2
1
Turn ON the main power supply switch with
the power cord connected.
2
Set the instrument to sleep mode.
PASS/FAIL
lamp
3
3
Press and hold the
STOP
button, and press
the start button.
4
Remove your hands when the
PASS/FAIL
lamp turns red.
5
5
Tap
[Reset all Settings.]
within 10 seconds.
Tap
[Exit]
when not implementing a full reset.
14
Maintenance and Service
Содержание ST4030
Страница 2: ......
Страница 14: ...6 Usage Notes ...
Страница 32: ...24 Inspecting before Measurement and Verifying Operations ...
Страница 42: ...34 Table Copying ...
Страница 108: ...100 Setting Dielectric Break down Judgment Conditions ...
Страница 132: ...124 Terminal Open Error Setting ...
Страница 198: ...190 Editing Files and Folders ...
Страница 214: ...206 Interface Specifications ...
Страница 237: ...229 Dimensions Diagram 15 4 Dimensions Diagram Unit mm 215 1 348 3 178 0 5 285 6 0 5 200 1 13 40 3 1 15 3 1 15 Appendix ...
Страница 248: ...240 Precautions when Processing Test Leads ...
Страница 252: ...244 Index ...
Страница 253: ......
Страница 254: ......
Страница 255: ......
Страница 256: ......