218
Error Display
Display *
1
EXT.I/O
ERR pin
Communications
VCAL status
Meaning
INFORMATION 1013
ON
1
Indicates that voltage calibration was canceled
by the instrument’s interlock function.
INFORMATION 1021
OFF
104
The workpiece’s entire response waveform can
be recorded by either reducing the sampling
frequency or increasing the number of data
points.
For some workpieces, you may encounter
this message even if you use the slowest
sampling frequency and the maximum number
of sampling data points. In such cases, choose
the setting that is best suited to the waveform.
INFORMATION 1022
OFF
105
The workpiece’s entire response waveform can
be recorded by either increasing the sampling
frequency or decreasing the number of data
points. This will allow the measurement time to
be reduced.
*1: If the voltage was calibrated from communications, the window is not displayed.
USB host error
Display
Meaning
Solution
• USB memory voltage
consumption exceeds 500 mA.
Use another USB memory.
• USB memory format is
incompatible with the
instrument.
Use another USB memory.
• File is corrupt and so cannot be
imported.
• File is not compatible with the
instrument and so cannot be
imported.
Check that the file and USB memory
are not damaged.
• Insufficient USB memory space. Use another USB memory, or
increase the available space.
Содержание ST4030
Страница 2: ......
Страница 14: ...6 Usage Notes ...
Страница 32: ...24 Inspecting before Measurement and Verifying Operations ...
Страница 42: ...34 Table Copying ...
Страница 108: ...100 Setting Dielectric Break down Judgment Conditions ...
Страница 132: ...124 Terminal Open Error Setting ...
Страница 198: ...190 Editing Files and Folders ...
Страница 214: ...206 Interface Specifications ...
Страница 237: ...229 Dimensions Diagram 15 4 Dimensions Diagram Unit mm 215 1 348 3 178 0 5 285 6 0 5 200 1 13 40 3 1 15 3 1 15 Appendix ...
Страница 248: ...240 Precautions when Processing Test Leads ...
Страница 252: ...244 Index ...
Страница 253: ......
Страница 254: ......
Страница 255: ......
Страница 256: ......