211
Troubleshooting
Communications
Symptom
Check item or cause
Corrective action and reference page
Communications are disabled
(
[REMOTE]
icon is not
displayed)
Have communications been
established?
Check that the connector connections
are correct.
Check that the interface settings are
correct.
Interface” (p. 145)
(USB) Install the drivers on the control
devices.
See: Application disc “Communication
Command Instruction Manual”
(RS-232C) Use a cross cable.
Interface” (p. 145)
(USB, RS-232C) Check the COM port
number of the control devices.
(RS-232C) Align the baud rates of the
instrument and the control devices.
Communications are disabled
(
[REMOTE]
icon is
displayed)
Do the communications settings
match?
When commands are not received,
check the control software delimiters.
(GP-IB) Check the settings of the
message terminators.
(GP-IB) Check that the address settings
are correct.
A communications error
occurred
Has a command error occurred?
Commands are mismatched
• Check the command spelling. Space is
x20H.
• Do not add a ? to commands that are
not queries.
• (RS-232C) Align the baud rates of the
instrument and the control devices.
See: Application disc “Communication
Command Instruction Manual”
Has an implementation error
occurred?
When command strings are correct, but
unable to be implemented, try using
a mode, etc., that can implement the
commands.
See: Application disc “Communication
Command Instruction Manual”
Have you checked the command
send and receive details using
the communications monitoring
function?
Check the command send and receive
details using the communications
monitoring function.
14
Maintenance and Service
Содержание ST4030
Страница 2: ......
Страница 14: ...6 Usage Notes ...
Страница 32: ...24 Inspecting before Measurement and Verifying Operations ...
Страница 42: ...34 Table Copying ...
Страница 108: ...100 Setting Dielectric Break down Judgment Conditions ...
Страница 132: ...124 Terminal Open Error Setting ...
Страница 198: ...190 Editing Files and Folders ...
Страница 214: ...206 Interface Specifications ...
Страница 237: ...229 Dimensions Diagram 15 4 Dimensions Diagram Unit mm 215 1 348 3 178 0 5 285 6 0 5 200 1 13 40 3 1 15 3 1 15 Appendix ...
Страница 248: ...240 Precautions when Processing Test Leads ...
Страница 252: ...244 Index ...
Страница 253: ......
Страница 254: ......
Страница 255: ......
Страница 256: ......