Discharge Judgments (When Incorporating Model ST9000)
Threshold value of the test waveform discharge amount over the
master waveform discharge amount
When the judgment method is
[FIXED]
(Measurement screen)
[MODE]
>
[SETTING]
>
[JUDGE]
>
[DISCHARGE]
2
1
3
1
Tap
[LIMIT]
.
2
Enter the threshold value of the test waveform discharge amount over the master waveform
discharge amount.
Settings range: 0% to 999% (resolution: 1)
OFF
Discharge amount operations are implemented, but judgments are not.
3
Tap
[SET]
to confirm.
IMPORTANT
• When the threshold value is 100%, the discharge amount has the same threshold value as the master
waveform discharge value.
• When the threshold value is 200%, the discharge amount has double the threshold value of the master
waveform discharge value.