76
Discharge Judgments (When Incorporating Model ST9000)
Setting the discharge amount judgment area
Sets the extend of the the response waveform for the discharge judgment to target.
Shorten the judgment area to reduce the operations time.
(Measurement screen)
[MODE]
>
[SETTING]
>
[JUDGE]
>
[DISCHARGE]
1
1
Set the end point for the judgment area.
Setting range: 1001 points to 8001 points (resolution: 1000 points)
Decreases by 1000.
Increases by 1000.
IMPORTANT
Set the settings range so that the vibration waveform of the self-resonant area includes at least 4 periods. If
the settings range is too narrow, the discharge judgment will always be FAIL.