95
Setting Dielectric Break down Judgment Conditions
7.4
Setting Dielectric Break down Judgment
Conditions
Sets the conditions for judging dielectric break down from variations in the LC and RC values
between multiple pulses, discharge amount, waveform surface area values, peak voltage values,
and vibration frequency for each applied voltage.
LC and RC values judgments
Judges breakdown from the variations in the LC and RC values between multiple pulses for each
applied voltage.
If the variations in the LC and RC values between multiple pulses are greater than the set deviation
threshold value, dielectric break down is judged to have occurred.
(Measurement screen)
[MODE]
>
[BDV]
>
[JUDGE]
1
2
1
Tap
[LCRC AREA]
.
2
Set the deviation threshold value.
Increases by 1.
Decreases by 1.
C
Sets to the default values.
OFF
LC and RC values judgments are unexecuted.
7
Break down voltage test (BDV)
Содержание ST4030
Страница 2: ......
Страница 14: ...6 Usage Notes ...
Страница 32: ...24 Inspecting before Measurement and Verifying Operations ...
Страница 42: ...34 Table Copying ...
Страница 108: ...100 Setting Dielectric Break down Judgment Conditions ...
Страница 132: ...124 Terminal Open Error Setting ...
Страница 198: ...190 Editing Files and Folders ...
Страница 214: ...206 Interface Specifications ...
Страница 237: ...229 Dimensions Diagram 15 4 Dimensions Diagram Unit mm 215 1 348 3 178 0 5 285 6 0 5 200 1 13 40 3 1 15 3 1 15 Appendix ...
Страница 248: ...240 Precautions when Processing Test Leads ...
Страница 252: ...244 Index ...
Страница 253: ......
Страница 254: ......
Страница 255: ......
Страница 256: ......