168
Z2010, Z2011 Humidity Sensor
Humidity
measurement
accuracy
±3% RH (20°C to 30°C , 20% RH to 90% RH)
See the figure below if the humidity is outside of the range.
Temperature [°C]
Relative humidity [% RH]
Values marked with an asterisk (*) are
outside the guaranteed accuracy range (reference value).
80
0
10
20
40
60
20
30
80
100
0
40
50
60
70
±8% RH*
±8% RH*
±8% RH*
±8% RH*
±8% RH
±6% RH
±6% RH
±5% RH
±5% RH
±5% RH
±5% RH
±5% RH
±4% RH
±4% RH
±6% RH
±6% RH
±6% RH*
±6% RH*
±10% RH*
±12% RH*
±12% RH*
±12% RH*
±6% RH*
±3% RH
±3.5% RH
Hysteresis: ±1% RH (Added to the humidity measurement accuracy)
Response time
Approx. 300 seconds
(Temperature: 90% response time for 0°C → 60°C, 60°C → 0°C
[reference value])
(Humidity: 90% response time for 10% RH → 80% RH, 80% RH →
10% RH [reference value])