76
Setting Measurement Conditions
CT9692-90
20 A, 200 A
CT9693-90
200 A, 2000 A
CT7631
10 A, 100 A
CT7636
20 A, 200 A
CT7642
200 A, 2000 A
CT7731
10 A, 100 A
CT7736
20 A, 200 A
CT7742
200 A, 2000 A
CT9667
500 A, 5000 A
CT7044
50 A, 500 A, 5000 A
CT7045
50 A, 500 A, 5000 A
CT7046
50 A, 500 A, 5000 A
4
Select the zero suppression setting.
Settings:
ON
Suppresses the value to zero when the count is less than 10 in each range.
OFF
Displays the value even when the count is less than 10 (for zero
adjustment for DC measurement).
5
Select the recording mode.
Settings:
Instantaneous
Records the instant value for each recording interval.
Average
Measures every 0.5 seconds and records the average value for each
recording interval. (When the recording interval is less than 1 second,
the instant value is recorded.)
Maximum*
Measures every 0.5 seconds and records the maximum value for each
recording interval. (When the recording interval is less than 1 second,
the instant value is recorded.)
* The Model LR8513 Wireless Clamp Logger firmware version 1.30 or later is required to
be installed to use the instrument with the recording mode setting set to Maximum.
6
Select the filter setting.
Settings:
OFF
Does not use the filter.
Current measurement including high-frequency components can be
performed.
ON
For 50 Hz/60 Hz current measurement, unnecessary high-frequency
components can be cut.
With the scaling function, the measured current value can be displayed in a power