![Hioki LR8512 Скачать руководство пользователя страница 159](http://html1.mh-extra.com/html/hioki/lr8512/lr8512_instruction-manual_2137140159.webp)
153
LR8513 Wireless Clamp Logger
Measurement range/
Allowable
measurement range/
Resolution
Measurement
object
Sensor
used
Range
Maximum
resolution
Allowable
measurement range
Current
CT7731
10.00 A
0.01 A
0.10 A to 20.00 A AC
±(0.10 A to 20.00 A) DC
100.0 A
0.1 A
1.0 A to 100.0 A AC
±(1.0 A to 100.0 A) DC
CT7736
20.00 A
0.01 A
0.10 A to 20.00 A AC
±(0.10 A to 20.00 A) DC
200.0 A
0.1 A
1.0 A to 200.0 A AC
±(1.0 A to 200.0 A) DC
CT7742
200.0 A
0.1 A
1.0 A to 200.0 A AC
±(1.0 A to 200.0 A) DC
2000 A
1 A
10 A to 2000 A AC
±(10 A to 2000 A) DC
CT9667-01
500.0 A
0.1 A
1.0 A to 500.0 A AC
5000 A
1 A
10 A to 5000 A AC
CT9667-02
500.0 A
0.1 A
1.0 A to 500.0 A AC
5000 A
1 A
10 A to 5000 A AC
CT9667-03
500.0 A
0.1 A
1.0 A to 500.0 A AC
5000 A
1 A
10 A to 5000 A AC
CT7044
50.00 A
0.01 A
0.10 A to 50.00 A AC
500.0 A
0.1 A
1.0 A to 500.0 A AC
5000 A
1 A
10 A to 5000 A AC
CT7045
50.00 A
0.01 A
0.10 A to 50.00 A AC
500.0 A
0.1 A
1.0 A to 500.0 A AC
5000 A
1 A
10 A to 5000 A AC
CT7046
50.00 A
0.01 A
0.10 A to 50.00 A AC
500.0 A
0.1 A
1.0 A to 500.0 A AC
5000 A
1 A
10 A to 5000 A AC
Measurement accuracy
±0.5% rdg. ±5 dgt. (DC, AC 50 Hz/60 Hz)
When zero suppress is set to OFF, any value less than 10 counts is
outside the guaranteed accuracy.
(When a clamp on sensor is connected, the accuracy of the clamp
on sensor is added.)
Any intermittent current and leak current cannot be measured.
Temperature
characteristics
Measurement accuracy × 0.05/°C (added to the measurement
accuracy when the temperature is outside 23 ± 5°C)
Filter
ON/OFF selection
Frequency
characteristics (-3 dB)
180 Hz ± 30 Hz (with the filter ON)
1 kHz (with the filter OFF)
RMS value calculation
True RMS value calculation by software
(The reversed current is indicated with a - symbol during DC
measurement.)
www.
GlobalTestSupply
.com
Find Quality Products Online at: