159
LR8514 Wireless Humidity Logger
Function
Alarm
Displays
[ALARM]
when the value is outside of the setting range.
Lit:
The current measurement value is outside the range.
Blinking: There was a value that was outside the range but the
current value is in the range.
Scaling
Scales and displays a measurement value.
(The lamp turns off if a unit other than the allowable display unit is
set during scaling.)
Recording operation
hold function
Automatically recovers records after the power is recovered if the
power is shut off during recording.
(The data during power shutdown will not be recorded.)
Erroneous operation
prevention function
Displays a confirmation message when recording starts/stops, or
the power is turned off.
Whether a confirmation message is displayed can be selected.
Comment recording
function
Records a comment in the title or each channel (up to 40 single-
byte characters)
(number, English character, symbol).
Power saving function
Energy can be saved by turning ON/OFF the Bluetooth power.
1. Manual ON/OFF using the power key
2. Automatic ON/OFF according to the schedule setting
Energy saving can be set according to the date, day, or time (10
minutes to 12 hours).
Authentication function
Authentication using a password (ON/OFF)
Free Run
ON/OFF selection
ON: The measurement value is indicated every 1 second while
measurement is stopped. (the data is not saved in the memory)
The measurement value is saved in the memory every
recording interval and indicated every 1 second regardless
of recording interval setting while measuring. (when the
setting of recording interval is less than 1 second, the
measurement value is indicated every recording interval)
OFF: The measurement value which is at the time of a
measurement stop is indicated while measurement is
stopped.
The measurement value is saved in the memory and
indicated every recording interval while measuring.