background image

75

Setting Measurement Conditions

LR8513 Wireless Clamp Logger

Making settings for current measurement

The current is measured.

4

3

2

5

6

1

ON

Select the channel to be set and check the Measurement checkbox 

[

]

 

(ON).

Select the clamp sensor to be used.

Settings:

9675

9657-10

9695-02

CT6500

9669

CT9691-90

,

CT9692-90

CT9693-90

CT7631*

2

CT7636*

2

CT7642*

2

CT7731*

2

CT7736*

2

CT7742*

2

CT9667*

1

*

2

CT7044*

2

CT7045*

2

CT7046*

2

*1: Select CT9667 for the CT9667-01, CT9667-02, and CT9667-03.

*2:  The Model LR8513 Wireless Clamp Logger firmware version 1.40 or later is required 

to be installed to use the following clamp sensors: CT7044, CT7045, CT7046, CT7631, 

CT7636, CT7642, CT7731, CT7736, CT7742, or CT9667.

Select the measurement range according to the object to be measured.

Settings:

9675

500 mA, 5 A

9657-10

500 mA, 5 A

9695-02

5 A, 50 A

CT6500

50 A, 500 A

9669

1000 A

CT9691-90

10 A, 100 A

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Содержание LR8512

Страница 1: ...SS PULSE LOGGER EN Sept 2018 Revised edition 6 LR8512B980 06 18 09H LR8514 LR8515 WIRELESS CLAMP LOGGER WIRELESS HUMIDITY LOGGER WIRELESS VOLTAGE TEMP LOGGER www GlobalTestSupply com Find Quality Prod...

Страница 2: ...aration for Measurements 29 2 1 Inspection Before Use 29 2 2 Inserting Replacing Batteries 30 Installation Replacement 31 Battery indicator display 32 Battery life indication 33 2 3 Connecting the AC...

Страница 3: ...ay Configuration 58 4 5 Basic Operation Procedure 59 Operation flow 59 Selecting multiple wireless loggers 60 4 6 Registering Deleting a Wireless Logger 61 Registering a wireless logger 61 Deleting a...

Страница 4: ...ent data 103 4 12 Performing Wireless Logger Maintenance 104 Setting the clock 104 Self diagnosis 105 Checking the version number 106 Updating the firmware version 107 4 13 Convenient Functions 108 5...

Страница 5: ...data collection 136 5 10 Browsing Analyzing Measurement Data 137 Browsing measurement data 137 Exporting measurement data 139 Deleting measurement data 142 5 11 Performing Wireless Logger Maintenance...

Страница 6: ...teries 174 Appendix Appx 1 Measurement value recording Appx 1 Recording interval and recording time Appx 4 Target collection time Appx 4 Initial setting list Appx 5 www GlobalTestSupply com Find Quali...

Страница 7: ...nuals for the intended usage Measurement Guide Read this first For customers who use this instrument for the first time this manual describes the basic operation procedures Instruction Manual this man...

Страница 8: ...r or reseller Check the package contents as follows Instrument LR8512 1 LR8513 1 LR8514 1 LR8515 1 Common accessories CD 1 Instruction Manual PDF Application Software Logger Utility Wireless Logger Co...

Страница 9: ...19 Connection Cable for 9695 02 connection CT7631 AC DC Current Sensor CT7636 AC DC Current Sensor CT7642 AC DC Current Sensor CT7731 AC DC Auto Zero Current Sensor CT7736 AC DC Auto Zero Current Sens...

Страница 10: ...gth including the sensor Approx 50 mm Z2011 Humidity Sensor Cable length Approx 1 5 m Supported instrument LR8410 Wireless Logging Station Supported for software version 1 30 and later www GlobalTestS...

Страница 11: ...y to the operator WARNING Indicates a potentially hazardous situation that may result in death or serious injury to the operator CAUTION Indicates a potentially hazardous situation that may result in...

Страница 12: ...hat the product conforms to regulations set out by the EU Directive Indicates that the product incorporates Bluetooth wireless technology FCC ID Indicates the ID number of the wireless module certifie...

Страница 13: ...han that for which the instrument is rated could result in a severe accident and must be carefully avoided Using a measuring instrument without categories in an environment designated with the CAT II...

Страница 14: ...between the instrument and the LR8410 is 30 m If there is an obstacle wall metal screen etc communications may become unstable or the communications distance may become shorter Communications between...

Страница 15: ...cifications CAUTION This instrument is not drip proof Water that drips on the connector could cause a malfunction Do not allow any condensation to form Condensation can form particularly in an environ...

Страница 16: ...connector To prevent any wires from breaking pull on the connector end not the cord to disconnect the power cord from the outlet or the instrument The cables become stiff below 0 C If the cables are b...

Страница 17: ...nics may not operate properly and the life of the operator may be put at great risk WARNING Swallowing magnets could be life threatening Keep any magnets out of reach especially of small children If y...

Страница 18: ...g to the instrument may damage the instrument When connecting an external power supply connect the output plug to the instrument and then supply external power Handling of batteries WARNING Do not sho...

Страница 19: ...ry level due to the battery characteristics settings during use temperature or consumption level When a battery is used in a low temperature environment or a weak and deteriorated battery is used the...

Страница 20: ...exceeds the specified rating is applied the semiconductor relay can fail with a short circuit Never input any voltage that exceeds the specified rating Especially be aware of lightning surges If ther...

Страница 21: ...be measured to prevent damage to the instrument or electric shocks Handling of CD Do not allow any dirt or scratches on the disk recording surface When writing on the label face use a pen with a soft...

Страница 22: ...mp sensor Clamp sensor 9669 Clamp On Sensor CAT III 600 V AC 9695 02 CAT III 300 V AC CT6500 CAT III 600 V AC 9657 10 Clamp On Leak Sensor CAT III 300 V AC 9675 CT9691 90 Clamp On AC DC Sensor CAT III...

Страница 23: ...42 2000 A CT9667 01 AC Flexible Current Sensor 10000 A CT9667 02 10000 A CT9667 03 10000 A CT7044 1000A 600A range 10000A 6000A range CT7045 CT7046 Supplying a current in excess of the maximum input m...

Страница 24: ...round cable is close to a high voltage live part and it is difficult to measure change the route of the ground cable before measurement When the 9657 10 9675 Clamp On Leak Sensor is used CAUTION To pr...

Страница 25: ...teriorate the sensor characteristics or cause an open close operation failure To avoid damaging the instrument do not connect any equipment other than the clamp sensor to the BNC terminal To avoid dam...

Страница 26: ...luene etc increasing the error component of humidity measurement Do not expose the temperature and humidity sensor to any concentrated chemical solvent for an extended period of time while it is used...

Страница 27: ...LR8410 firmware version 1 30 or later The firmware version for the LR8410 is displayed on the system screen The latest version can be downloaded from our website For details on how to upgrade the sof...

Страница 28: ...needs to be updated The software can be updated in Wireless Logger Collector Windows PC version p 107 If the software version is older than 1 20 a communication error protocol error occurs in any comm...

Страница 29: ...Logger Utility LAN USB communications Windows PC Wireless Logger Collector 0 1 sec high speed sampling The data update of the LR8513 and LR8514 is every 0 5 sec The memory capacity for each channel is...

Страница 30: ...from the left indicate the year of manufacture and the next two indicate the month of manufacture It is necessary for production control requirements such as the product warranty Do not peel off the l...

Страница 31: ...toring MAX MIN AVG Maximum value Minimum value Average value Data number Unit number Date and time Operation verification 2 Measurement Lit Performing measurement Blinking Waiting for a measurement st...

Страница 32: ...or which the scaling setting is ON is not displayed If it matches the unit for which a segment is prepared however the unit is displayed 3 It is displayed only when connected to the LR8410 4 For pulse...

Страница 33: ...instruction manual 10 When scaling is set to ON the allowable display range is indicated in the following table Scaling displayed digit setting U F display Underflow Allowable display range O F displ...

Страница 34: ...28 Display Configuration Example www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Страница 35: ...instrument and clamp sensor connection cable If it is damaged there is a risk of electric shocks Do not use the instrument but instead send it for repair Does the display turn on when the power is tu...

Страница 36: ...life varies depending on the capacity charging condition deterioration due to repeated use etc Use batteries in consideration of the factors above Use of manganese dioxide batteries Do not use mangan...

Страница 37: ...tery 2 Phillips screwdriver No 2 1 Remove the connection cables 5 3 2 7 4 6 2 Using a Phillips screwdriver remove the screw 1 location from the battery cover The screw cannot be removed from the batte...

Страница 38: ...as possible Blinks The battery is drained Recording or communications cannot be made in this condition Off The battery indicator display turns off when the AC adapter or an external power supply is c...

Страница 39: ...in Real time measurement Bluetooth ON Approx 5 days Approx 7 days Approx 10 days Approx 14 days Manual data collection Bluetooth ON Approx 10 days Approx 10 days Approx 20 days Approx 20 days Manual d...

Страница 40: ...oth ON Approx 2 days Approx 4 days Approx 7 days Approx 14 days Manual data collection Bluetooth ON Approx 3 days Approx 7 days Approx 14 days Approx 20 days Manual data collection Bluetooth OFF Appro...

Страница 41: ...0 Hz AC adapter Power cord 3 1 2 1 Connect the power cord and AC adapter 2 Connect the outlet plug to the instrument AC adapter connection terminal left side 3 Connect the output plug to an outlet Whe...

Страница 42: ...s attached to the instrument the magnet can be attached to the wall surface metal plate 1 2 3 1 Attach the Z5004 or Z5020 through the strap hole In order of 1 2 3 Magnet 2 Attach it to the wall surfac...

Страница 43: ...ength approx 1 5 m Not used If a wrong input on the instrument side or connection destination is made or the L1010 is not completely inserted a correct value is not displayed If a correct value is not...

Страница 44: ...t to lock it When removing the connector turn it to the left and pull it out 3 Clamp the object to be measured p 43 If it is not completely inserted a correct value is not displayed If a correct value...

Страница 45: ...Flexible Current Sensor CT9667 02 CT9667 03 CT7044 2 CT7045 2 CT7046 2 1 The 9219 Connection Cord is required in order to connect a compatible clamp sensor 2 The CM7290 or CM7291 Display Unit and L90...

Страница 46: ...LED Green lamp lit Power ON Red lamp lit Low battery Lamp off Power OFF AC adapter connection terminal Sensor connection terminal Connect the clamp sensor Battery cover Remove the screw and insert bat...

Страница 47: ...mV is output for 10 A when the L range is selected on the CT9691 The voltage corresponding to the input current is output For example 100 mV is output for 10 A when the L range is selected on the CT96...

Страница 48: ...Select the output mode 1 6 2 6 3 5 1 Connect the current sensor to the CM7290 CM7291 2 Supply power to the CM7290 CM7291 The CM7290 CM7291 can operate on AC adapter or battery power 3 Connect output...

Страница 49: ...1 and the LR8513 may differ due to differences in the devices accuracy When measuring a negative DC current in AC DC measurement mode the CM7290 CM7291 and LR8513 may indicate different polarity readi...

Страница 50: ...mp On AC DC Sensor The figure is for the CT6500 9675 9657 10 Clamp On Leak Sensor The figure is for the 9675 OK NO OK NO OK NO OK NO For details refer to the instruction manual for the clamp sensor to...

Страница 51: ...easurement Clamp 3 wires at a time Clamp 3 wires at a time Clamp all the wires at the center For a single phase 2 wire circuit clamp 2 wires For a three phase 4 wire circuit clamp 4 wires When the wir...

Страница 52: ...uipment Leak current lg Class B ground cable Insulation deteriorated area A 2 If any short circuit is found inspect from the power supply to the load using batch measurement circuit clamped all at a t...

Страница 53: ...ge 9669 1000 A range 9675 Clamp On Leak Sensor 5 A range 9657 10 5 A range CT9691 90 Clamp On AC DC Sensor 100 A range CT9692 90 200 A range CT9693 90 2000 A range CT7631 AC DC Current Sensor 100 A ra...

Страница 54: ...en the sensor is completely inserted the instrument or the sensor may have a problem Send the instrument for repair Connecting the input cable thermocouple LR8515 Attach input cables or thermocouples...

Страница 55: ...nt Using the LR8410 1 Turn ON the power of the instrument and the LR8410 2 Register the instrument as a unit using the LR8410 Quick Set function Unit Guide 3 Set the measurement conditions using the Q...

Страница 56: ...rt up Logger Utility The main screen is displayed when Logger Utility starts up The main screen is the basic screen for Logger Utility 4 Set the measurement conditions for the instrument 5 Send the me...

Страница 57: ...r more 65 536 colors or more Recommended Use a small font When a large font is used the display may be disturbed Hard disk 3 0 GB or more of available space OS Windows 10 32 bit 64 bit Windows 8 32 bi...

Страница 58: ...ata can be collected repeatedly and periodically 10 minutes to 1 day Analysis browsing Measurement data can be displayed in the waveform format and analyzed using the Logger Utility Measurement data o...

Страница 59: ...nits p 61 When no wireless logger is found When the symbol in the screen is off press the power key to turn ON the Bluetooth function When the symbol in the screen is off a wireless connection is not...

Страница 60: ...ment data can be collected during measurement and after measurement stop 8 Analyze the data on the Windows PC 1 Display measurement data in the Logger Utility The data is displayed in a graph and anal...

Страница 61: ...g make sure to close it before starting the installation If anti virus software is running the software may not be installed properly 2 Set the provided CD in the CD ROM drive The installer starts aut...

Страница 62: ...lay the Apps view and click HIOKI Wireless Logger Collector For Windows 10 From the Windows Start menu click All Apps HIOKI Wireless Logger Collector Termination procedure In the main screen click the...

Страница 63: ...click Windows System Control Panel For Windows 10 Right click the Windows Start button and then click Control Panel 2 Click Programs and Features The Uninstall or change a program screen is displayed...

Страница 64: ...ess logger list tree display Switches the group display and list display Creates and deletes groups 6 Title display Displays the information on the wireless logger that is currently selected 7 Wireles...

Страница 65: ...e Maintenance 2 Select a group and wireless logger from the wireless logger list tree display 3 Select the item to be displayed from the wireless logger channel file list list display 4 Press the oper...

Страница 66: ...g on the communications with the wireless logger Select all Click Select All on the Ribbon bar Unselect Click Unselect on the Ribbon bar Select individual Click a checkbox in the list Select multiple...

Страница 67: ...he PC specifications Communications may not be established depending on the combination of the Bluetooth module and Bluetooth stack device driver Preparation 1 Place the wireless logger near the Windo...

Страница 68: ...gers that have been connected are displayed in the list on the left COM ports are displayed if a third party Bluetooth stack is used 3 Select the wireless logger s to be registered from the list on th...

Страница 69: ...the wireless loggers When an unauthorized virtual COM is registered the operation of Wireless Logger Collector may become unstable Ways of performing a pairing and creating a virtual serial port vary...

Страница 70: ...is displayed 2 Select the wireless logger s to be deleted from the list on the right multiple units can be selected 3 Press to delete the selected wireless logger s The selected wireless logger s move...

Страница 71: ...ggers Groups can be added to the layered structure Procedure to create a group 2 1 1 From List the wireless logger list tree display select Computer or the group for which a new group is to be created...

Страница 72: ...he left key of the mouse and drop it on the target group Drag and drop Procedure to delete a group 2 1 1 From List the wireless logger list tree display select the group to be deleted 2 Press If wirel...

Страница 73: ...ement conditions for wireless loggers 2 1 3 1 Select Register Settings from the Navigation bar 2 Select the target wireless logger 3 Click Open on the Ribbon bar The settings dialog box is displayed w...

Страница 74: ...set to 0 1 seconds when disconnection detection is set to ON Example When the recording interval is set to 1 second Peak 1 second Continuous Recording Allows you to set a processing method when the me...

Страница 75: ...reset start function is disabled Starts recording at the preset time The instrument enters the recording start standby state at the same time as the Scheduled Record Start ON setting is sent Measureme...

Страница 76: ...to ON OFF Settings Does not perform measurement Performs measurement Comment Allows you to set the channel comment Up to 40 single byte characters The characters entered are converted to the following...

Страница 77: ...pulses is reset for each recording interval Add Measures the number of integrated pulses after measurement starts Addition Instantaneous HIGH LOW Detection level 3 Select the measurement reference sl...

Страница 78: ...ommended to perform measurement with the input type Instantaneous and add the measurement values using Microsoft Excel later Making settings for rotation speed measurement The number of pulses output...

Страница 79: ...l Settings 1 V Determines 1 0 V or higher to be HIGH 0 V to 0 5 V to be LOW 4 V Determines 4 0 V or higher to be HIGH 0 V to 1 5 V to be LOW 6 Select the chattering prevention filter setting Settings...

Страница 80: ...1 V Determines 1 0 V or higher to be HIGH 0 V to 0 5 V to be LOW 4 V Determines 4 0 V or higher to be HIGH 0 V to 1 5 V to be LOW 4 Select the chattering prevention filter setting Settings OFF Turns O...

Страница 81: ...2 CT9667 1 2 CT7044 2 CT7045 2 CT7046 2 1 Select CT9667 for the CT9667 01 CT9667 02 and CT9667 03 2 The Model LR8513 Wireless Clamp Logger firmware version 1 40 or later is required to be installed to...

Страница 82: ...average value for each recording interval When the recording interval is less than 1 second the instant value is recorded Maximum Measures every 0 5 seconds and records the maximum value for each reco...

Страница 83: ...kbox ON Disconnection Detection BURN OUT is displayed for the measurement value and BURN is displayed on the screen under the following conditions When the temperature and humidity sensor is not conne...

Страница 84: ...surement range according to the object to be measured Settings 50 mV 500 mV 5 V 50 V When measuring the output from a measurement instrument When inputting a 4 mA to 20 mA current connect a 250 resist...

Страница 85: ...hermocouple type Settings K T 4 Select the disconnection detection setting Settings OFF Does not perform disconnection detection Normally when the thermocouple is disconnected a measurement value may...

Страница 86: ...e temperature measurement accuracy only Disconnection detection When disconnection detection is set to ON a fine current is input at the specified recording interval during thermocouple measurement an...

Страница 87: ...ttings OFF No scaling Decimal Displays the converted value as a decimal number Exponential Displays the converted value as an index number Wireless logger display is in a decimal number only 2 Select...

Страница 88: ...of the unit to be converted AH AL VH VL V Unit eu Example Actual measurement value input value With the scaling function the current value measured by the LR8513 Wireless Clamp Logger can be measured...

Страница 89: ...ling Decimal Displays the converted value as a decimal number Exponential Displays the converted value as an index number Wireless logger display is in a decimal number only 2 Set the unit to be conve...

Страница 90: ...tings AUTO Displays a 4 digit value 0 000 to 9 999 The decimal point position is changed as needed 0 Digit 1 Digit 2 Digit 3 Digit The decimal point is fixed to the specified digit Setting the auxilia...

Страница 91: ...r not to hold the alarm is set Settings OFF Cancels the alarm when the alarm conditions are no longer met ON Holds the alarm Composite Condition Performs logic synthesis for the alarm between channels...

Страница 92: ...dges it to be an alarm Window Out When a value is outside of the specified upper and lower limit range judges it to be an alarm Alarm Level HIGH LOW When Alarm Method is set to Level Settings HIGH Whe...

Страница 93: ...t between channels are met judges it to be an alarm AND When all of the conditions that are set between channels are met judges it to be an alarm X Ignores the signal 0 When the signal is LOW judges i...

Страница 94: ...odule power at the specified time on the specified date 1st to 31st When the specified date does not exist it is ignored Starting Communication Reception The clock time to turn ON the Bluetooth module...

Страница 95: ...for 10 seconds Start Backup Recording can be resumed when the power is lost due to a power outage etc during recording operation and the power is turned ON again Settings Does not use the start backu...

Страница 96: ...ording interval is less than 1 second the measurement value is indicated every recording interval Settings Does not use the free running function Use the free running function When the free run functi...

Страница 97: ...ice is specified communications cannot be established Security The authentication function using a password can be enabled to prevent the wireless logger from being controlled remotely without due car...

Страница 98: ...ettings cannot be copied if the wireless logger type is different 3 1 2 4 5 1 Select Register Settings from the Navigation bar 2 Select the wireless logger or channel from which the settings are copie...

Страница 99: ...tings The settings are restored to the default values p Appx 5 3 1 2 1 Select Register Settings from the Navigation bar 2 Select the target wireless logger 3 Click Clear www GlobalTestSupply com Find...

Страница 100: ...etting conditions edited in Wireless Logger Collector are sent to and set for wireless loggers using wireless communications 1 2 3 1 Select Register Settings from the Navigation bar 2 Select the targe...

Страница 101: ...eless communications 3 1 2 1 Select Register Settings from the Navigation bar 2 Select the target wireless logger 3 Press Receive When the settings are received the settings of the corresponding unit...

Страница 102: ...Wireless logger measurement is started and stopped 2 1 3 1 Select Measurement Monitoring from the Navigation bar 2 Select the target wireless logger 3 Press Start or Stop When measurement start is pr...

Страница 103: ...the monitoring interval is set p 110 Items to be monitored Measurement status latest measurement data battery level wireless signal level The monitoring result is displayed in each line of the list 2...

Страница 104: ...l AC adapter is used Wireless signal level Monitoring value Updated approximately once every second 4 When the selected wireless logger is being measured the latest measurement data is displayed When...

Страница 105: ...ven while wireless loggers are being measured Measurement data is periodically collected until the data collection is stopped once the collection interval is set p 110 2 1 3 1 Select Collection Browse...

Страница 106: ...he Navigation bar 2 Select the target wireless logger 3 Double click the target wireless logger in the list using the mouse The list switches to the collected measurement data file list 4 Select the m...

Страница 107: ...Navigation bar 2 Select the target wireless logger 3 Double click the target wireless logger in the list using the mouse The list switches to the collected measurement data file list 4 Select the mea...

Страница 108: ...Sections Time Format yyyy mm dd HH MM SS dd mm yyy HH MM SS mm dd yyyy HH MM SS Text Format Exponential Decimal 8 Select the location where the settings are to be saved Press Browse and select the de...

Страница 109: ...rget wireless logger 3 Double click the target wireless logger in the list using the mouse The list switches to the collected measurement data file list 4 Select the measurement data to be deleted fro...

Страница 110: ...ogger clock is set according to the Windows PC clock 2 1 3 1 Select Maintenance from the Navigation bar 2 Select the target wireless logger 3 Press Adjust Clock The clock cannot be set during wireless...

Страница 111: ...ation bar 2 Select the target wireless logger 3 Press Self Test Self diagnosis takes approximately 2 minutes If the self diagnosis result is displayed as Error the instrument needs to be repaired or i...

Страница 112: ...rsion number The firmware version of wireless loggers is checked 2 1 3 1 Select Maintenance from the Navigation bar 2 Select the target wireless logger 3 Press Version Check www GlobalTestSupply com F...

Страница 113: ...oaded from our Internet website Execute Import Version Upgrade File in the menu to load the downloaded file to Wireless Logger Collector CAUTION Never turn OFF the power of the wireless logger that is...

Страница 114: ...n the list can be changed Settings Small Standard Medium Large Extra Large Language The display language for the application is switched Restart the application to reflect the change Settings Japanese...

Страница 115: ...an be used when the Windows PC is transferred or backed up The registered wireless logger information and collected measurement data can be output all together as one file Import Application Data File...

Страница 116: ...s 1 to 7 device s The number of simultaneous connections may be limited depending on the PC environment Monitor Interval Settings Default settings Once 10 minutes 30 minutes 1 hour 2 hours 6 hours 12...

Страница 117: ...ment control Measurement can be started and stopped Monitoring function Numerical values can be monitored Update interval 1 second Measurement data collection function Measurement data can be download...

Страница 118: ...r Collector on the Android terminal p 115 Wireless Logger Collector 3 Register the instrument in Wireless Logger Collector up to 100 units p 122 4 Set the measurement conditions in Wireless Logger Col...

Страница 119: ...asurement p 133 You can also start measurement by holding down this button 2 3 1 6 Collect measurement data using Wireless Logger Collector p 136 2 3 1 www GlobalTestSupply com Find Quality Products O...

Страница 120: ...rt the measurement data to a file p 139 2 1 3 4 5 8 Copy the exported file onto the Windows PC p 140 USB cable 9 Analyze the data on the Windows PC www GlobalTestSupply com Find Quality Products Onlin...

Страница 121: ...software according to the instructions displayed in the screen Tap Play Store in the application screen Play Store may not be available on some models Tap the Search button Enter a key word Wireless...

Страница 122: ...from Google Play Store or the Settings Apps screen IMPORTANT When the application is uninstalled the wireless logger registration information settings and collected measurement data are erased Export...

Страница 123: ...ation buttons Settings Measurement Collection Browse Menu buttons Registered wireless loggers Group name Wireless logger monitoring status Wireless logger search and registration Back www GlobalTestSu...

Страница 124: ...onfiguration Monitor screen Latest measurement data Trend waveform Update interval 1 second up to 100 data Wireless logger status www GlobalTestSupply com Find Quality Products Online at sales GlobalT...

Страница 125: ...settings to the selected wireless logger Receives all settings from the selected wireless logger Confirms and saves the changed settings Discards the changed settings www GlobalTestSupply com Find Qu...

Страница 126: ...isplay Configuration Browse screen Collection date and time Title comment Number of data Measurement start date and time www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply...

Страница 127: ...the specified time Displayed waveforms can be expanded or compressed along the horizontal axis Select channel for gauge display The display range can be adjusted www GlobalTestSupply com Find Quality...

Страница 128: ...egistered 1 2 1 Tap The search result list is displayed It may take several tens of seconds to search for wireless loggers The wireless loggers that are located nearby and are not registered are displ...

Страница 129: ...e Logger 3 Select the group to which the unit to be deleted belongs 4 Select the unit to be deleted IMPORTANT When wireless loggers are deleted registration canceled the settings and collected measure...

Страница 130: ...managed Grouping is useful for managing multiple wireless loggers Procedure to create a group 1 2 1 Tap the menu button 2 Tap Add Group 3 Enter a new group name A new group is now created at the very...

Страница 131: ...ogger to be moved 5 Select the group to which the wireless logger is to be moved Procedure to delete a group 1 2 1 Tap the menu button 2 Tap Remove Group 3 Select the group to be deleted If wireless l...

Страница 132: ...gers For details on settings see 4 7 Setting Measurement Conditions p 67 1 2 3 1 Tap Settings button 2 Tap the wireless logger for which the settings are to be changed The settings screen is displayed...

Страница 133: ...aracters Identifier Name You can name the unit to be identified up to 16 single byte characters Recording Interval Allows you to set the interval to import data Continuous Recording Allows you to set...

Страница 134: ...onditions Setting the channel Set the input scaling and alarm for each channel Alarm settings screen Scaling settings screen Settings screen www GlobalTestSupply com Find Quality Products Online at sa...

Страница 135: ...ooth module power is ON is set Security An authentication function using a password can be enabled to prevent the unit from being controlled carelessly by a third party When changing the password set...

Страница 136: ...ed every 1 second regardless of recording interval setting while measuring when the setting of recording interval is less than 1 second the measurement value is indicated every recording interval Init...

Страница 137: ...ations are sent to and set for wireless loggers using wireless communications 2 1 3 1 Tap the Settings button 2 Tap the wireless logger to be edited The settings screen is displayed 3 Tap the Send but...

Страница 138: ...lication using wireless communications 2 1 3 1 Tap the Settings button 2 Tap the wireless logger to which the settings are to be sent The settings screen is displayed 3 Tap the Receive button The meas...

Страница 139: ...urement Wireless logger measurement is started and stopped 2 3 1 1 Tap the Measure button 2 Tap the wireless logger for which measurement is to be started or stopped The menu is displayed 3 Select Sta...

Страница 140: ...status of all the registered wireless loggers is acquired only once Item Icon Description Power AC Battery level 1 It is recommended to replace the batteries Battery level 2 Battery level 3 Wireless s...

Страница 141: ...the wireless logger to be monitored The menu is displayed 3 Tap Start Monitoring in the menu The monitoring screen is displayed p 118 The monitoring value is updated every second When the selected wi...

Страница 142: ...eless loggers Measurement data can also be collected during measurement 2 3 1 1 Tap the Collection button 2 Tap the wireless logger from which the measurement data is to be collected The menu is displ...

Страница 143: ...ment Data 5 10 Browsing Analyzing Measurement Data Browsing measurement data The collected measurement data can be displayed on the screen 2 1 3 4 www GlobalTestSupply com Find Quality Products Online...

Страница 144: ...reless logger for which the measurement data is to be browsed The browsing screen is displayed p 120 3 Tap the collected measurement data to be displayed The menu is displayed 4 Tap Show Graph www Glo...

Страница 145: ...the Windows PC etc 2 1 3 4 5 1 Tap the Browse button 2 Tap the wireless logger for which the measurement data is to be browsed The browsing screen is displayed 3 Tap the collected measurement data to...

Страница 146: ...tion A file name is automatically generated HIOKI Wireless Logger Collector Model number and serial number Copying the output file onto the Windows PC When the file is output to the internal storage o...

Страница 147: ...ze the file using the supplied Logger Utility When the file is output to an external medium SD card USB flash drive etc 1 Remove the medium from the Android terminal and connect it to the Windows PC T...

Страница 148: ...the wireless logger for which the measurement data is to be deleted The browsing screen is displayed 3 Tap the measurement data to be deleted The menu is displayed 4 Tap Delete Data IMPORTANT Be awar...

Страница 149: ...Tap the menu button 2 Tap Maintenance The system enters the maintenance mode 3 Select the wireless logger for which maintenance is to be performed The menu is displayed 4 Select Adjust Clock or Versio...

Страница 150: ...ck is set according to the Android terminal clock The clock cannot be set during wireless logger measurement and monitoring Checking the version number The wireless logger version number is displayed...

Страница 151: ...Auto calendar automatic leap year adjustment 24 hour clock Clock accuracy 50 ppm reference value at temperature of 23 C 4 32 seconds per day Recording interval 0 1 sec 0 2 sec 0 5 sec 1 sec 2 sec 5 se...

Страница 152: ...OFF Rejects Bluetooth connection Operating environment Indoors Pollution Degree 2 altitude up to 2 000 m 6 562 ft Operating temperature and humidity Temperature 20 C to 60 C 4 0 F to 140 0 F Humidity...

Страница 153: ...instant Revolution Logic records 1 0 for each recording interval Pulse input Measurement range resolution Measurement object Range Maximum resolution Measurement range Integrating 1 000 mega pulses f...

Страница 154: ...not be recorded Erroneous operation prevention function Displays a confirmation message when recording starts stops or the power is turned off Whether a confirmation message is displayed can be select...

Страница 155: ...uetooth ON 0 05 VA with Bluetooth OFF Maximum rated power When the recording interval is 0 1 seconds during real time measurement Z2003 AC Adapter 4 5 VA AC adapter included 0 45 VA instrument only 13...

Страница 156: ...e value at temperature of 23 C 4 32 seconds per day Recording interval 0 5 sec 1 sec 2 sec 5 sec 10 sec 20 sec 30 sec 1 min 2 min 5 min 10 min 20 min 30 min 1 hour Recording mode Instant value recordi...

Страница 157: ...tting Bluetooth standby can be turned ON OFF by Power key operation ON Accepts Bluetooth connection OFF Rejects Bluetooth connection Operating environment Indoors Pollution Degree 2 altitude up to 2 0...

Страница 158: ...000 A AC 9695 02 5 000 A 0 001 A 0 010 A to 5 000 A AC 50 00 A 0 01 A 0 10 A to 50 00 A AC CT6500 50 00 A 0 01 A 0 10 A to 50 00 A AC 500 0 A 0 1 A 1 0 A to 500 0 A AC 9669 1000 A 1 A 10 A to 1000 A...

Страница 159: ...A to 500 0 A AC 5000 A 1 A 10 A to 5000 A AC CT7045 50 00 A 0 01 A 0 10 A to 50 00 A AC 500 0 A 0 1 A 1 0 A to 500 0 A AC 5000 A 1 A 10 A to 5000 A AC CT7046 50 00 A 0 01 A 0 10 A to 50 00 A AC 500 0...

Страница 160: ...tion Records a comment in the title or each channel up to 40 one byte characters number alphabet symbol Energy saving function Energy can be saved by turning ON OFF the Bluetooth power 1 Manual ON OFF...

Страница 161: ...A AC adapter included 0 45 VA instrument only 13 5 V DC external power supply 0 5 VA When batteries are used 0 3 VA Continuous operating time Battery When the recording mode and free run are set to in...

Страница 162: ...lock Clock accuracy 50 ppm reference value at temperature of 23 C 4 32 seconds per day Recording interval 0 5 sec 1 sec 2 sec 5 sec 10 sec 20 sec 30 sec 1 min 2 min 5 min 10 min 20 min 30 min 1 hour R...

Страница 163: ...ity 80 RH or less non condensation The operating temperature and humidity depend on the battery specifications when batteries are used Storage temperature and humidity Temperature 20 C to 60 C 4 0 F t...

Страница 164: ...C to 80 C The measurement accuracy depends only on the sensor Humidity measurement accuracy Z2010 Z2011 used 3 RH 20 C to 30 C 20 RH to 90 RH See the figure below if the humidity is outside of the ra...

Страница 165: ...racters number English character symbol Power saving function Energy can be saved by turning ON OFF the Bluetooth power 1 Manual ON OFF using the power key 2 Automatic ON OFF according to the schedule...

Страница 166: ...seconds during real time measurement Z2003 AC Adapter 4 5 VA AC adapter included 0 45 VA instrument only 13 5 V DC external power supply 0 5 VA Batteries 0 3 VA Continuous operating time Battery Appro...

Страница 167: ...ofile SPP Clock function Auto calendar automatic leap year adjustment and 24 hour clock Clock accuracy 50 ppm reference value at temperature of 23 C 4 32 seconds per day Recording interval 0 1 sec 0 2...

Страница 168: ...onnection OFF Rejects Bluetooth connection Operating environment Indoors Pollution Degree 2 altitude up to 2 000 m 6 562 ft Operating temperature and humidity Temperature 20 C to 60 C 4 0 F to 140 0 F...

Страница 169: ...tage 50 mV f s 0 01 mV 50 mV to 50 mV 0 05 mV 500 mV f s 0 1 mV 500 mV to 500 mV 0 5 mV 5 V f s 1 mV 5 V to 5 V 5 mV 50 V f s 10 mV 50 V to 50 V 50 mV Thermocouples Reference junction compensation acc...

Страница 170: ...input Recording interval of 1 second digital filter of 60 Hz 50 mV f s range for 60 Hz input Maximum input voltage 50 V DC Maximum rated voltage between input channels 60 V DC Maximum rated voltage to...

Страница 171: ...ding interval OFF The measurement value which is at the time of a measurement stop is indicated while measurement is stopped The measurement value is saved in the memory and indicated every recording...

Страница 172: ...FF 23 C Approx 7 days reference value at recording interval of 1 second Bluetooth ON 23 C Approx 2 days reference value at recording interval of 0 1 seconds during real time measurement 23 C When the...

Страница 173: ...nsation Temperature and humidity for guaranteed accuracy At temperature measurement Temperature 40 C to 80 C 40 0 F to 176 0 F Only the operating temperature and humidity of the logger when Z2010 is u...

Страница 174: ...10 20 40 60 20 30 80 100 0 40 50 60 70 8 RH 8 RH 8 RH 8 RH 8 RH 6 RH 6 RH 5 RH 5 RH 5 RH 5 RH 5 RH 4 RH 4 RH 6 RH 6 RH 6 RH 6 RH 10 RH 12 RH 12 RH 12 RH 6 RH 3 RH 3 5 RH Hysteresis 1 RH Added to the...

Страница 175: ...s of the instrument or installation environment We recommend that the calibration frequency is determined in accordance with the status of the instrument or installation environment and that you reque...

Страница 176: ...al packing materials that were used when the instrument was delivered Cleaning To clean the instrument wipe it gently with a soft cloth moistened with water or mild detergent When the clamp core butt...

Страница 177: ...ot work The instrument has an internal error or the keys have failed Remove the AC adapter and batteries and then turn on the instrument power If the problem persists the instrument needs to be repair...

Страница 178: ...er terminal LR8410 PC or Android terminal is connected is lit Cut off the communications of the connected terminal and then search for or connect the wireless logger again Windows non standard Bluetoo...

Страница 179: ...dware error An error has occurred in the hardware Err 140 Backup battery error An error has occurred during clock backup O F U F The measurement value is outside of the measurement range or display ra...

Страница 180: ...atteries Removing lithium batteries Required items 1 Phillips screwdriver No 2 and nippers 1 Remove the cables and power cord 2 Remove the screw that secures the bottom case on the instrument rear sid...

Страница 181: ...of integrated pulses for 2 seconds is P20 2 000 Revolution rt 2 for t 2 s is calculated as follows rt 2 2000 1000 4 250 r s Exception When time t s is less than 1 second When time t s is less than 1 s...

Страница 182: ...lculation An RMS value is calculated from the data of a maximum of 400 points when a measurement signal is sampled approximately every 240 s Maximum of 400 up to a pause of the period Approx 240 s The...

Страница 183: ...ding After recording starts the average maximum value in the recording interval is recorded As initial data the instant value immediately after the recording starts is recorded Recording interval Reco...

Страница 184: ...minutes 20 seconds 20 seconds 115 days 17 hours 46 minutes 40 seconds 30 seconds 173 days 14 hours 40 minutes 00 second 1 minutes 347 days 5 hours 20 minutes 00 second 2 minutes to 1 hour 365 days or...

Страница 185: ...LR8515 Range 1000 M LR8512 500 mA LR8513 100 C 100 RH LR8514 50 mV LR8515 Sensor 9675 LR8513 K thermocouple LR8515 REC Mode Instantaneous LR8513 Chattering Prevention Filter OFF LR8512 Filter OFF LR8...

Отзывы: