72
Setting Measurement Conditions
5
Select the chattering prevention filter setting.
Settings:
OFF
Turns OFF the chattering prevention filter.
ON
Turns ON the chattering prevention filter.
For mechanical contact (relay) output signals, a count error due to
chattering can be prevented.
•
With the scaling function, the number of integrated pulses can be converted to a
physical quantity (Wh, VA, etc.) for the object to be measured. (p. 81)
• If there is a possibility that the number of pulses may exceed 1,000,000,000, it is
recommended to perform measurement with the input type “Instantaneous” and
add the measurement values using Microsoft
®
Excel
®
later.
Making settings for rotation speed measurement
The number of pulses output according to the rotation speed of the rotary encoder or
speed recorder is measured. The number of input pulses for one second is counted
and the rotation speed is calculated.
1
3
4
2
5
6
:
ON
1
Select the channel to be set and check the Measurement checkbox
[
]
(ON).
2
Select
[Rotation]
.
3
Set the number of pulses per rotation output from the sensor.
Settings:
1
to
1,000
The range is automatically set according to the number of pulses.