77
―――――――――――――――――――――――――――――――――――――――
6.7 Numerical Calculation
―――――――――――――――――――――――――――――――――――
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
APP
IND
The mean value
Details of the various calculations
(1) Maximum value (Max)
・
The maximum value of the waveform data is shown (in
volts).
Time to maximum value (Max-Time)
・
The time period from the moment that triggering
occurred till the maximum value is shown (in seconds).
・
If the maximum value is attained twice or more in the
data, the first point is took as the maximum value.
(2) Minimum value (Min)
・
The minimum value of the waveform data is shown (in
volts).
Time to minimum value (Min-Time)
・
The time period from the moment that triggering
occurred till the minimum value is attained is shown (in
seconds).
・
If the minimum value is attained twice or more in the
data, the first point is took as the minimum value.
(3) Peak-to-peak value
The peak-to-peak value (the difference between the
maximum value and the minimum value) of the waveform
data is shown (in volts).
(4) Mean value
・
The average value (in volts) of the waveform data is
shown.
・
The equation used for the calculation is as follows:
n: number of data samples
d
i
: the ith data value of the source channel
Содержание 8804
Страница 1: ...MEMORY HiCORDER INSTRUCTION MANUAL 8804 ...
Страница 2: ......
Страница 28: ...6 1 2 System Operation ...
Страница 40: ...18 2 1 Main Units Specifications ...
Страница 90: ...68 6 3 Basic Operational Procedures 2 Press the PRINT key The manual print as the figure below is recorded ...
Страница 102: ...80 6 8 Memory Recorder Function Possible Recording Time ...
Страница 128: ...106 8 3 Basic Operational Procedures ...
Страница 200: ...178 12 4 Settings on the System Screen Page 2 ...
Страница 230: ...APPENDIX 8 Appendices Single Dual 2 Recorder function ...
Страница 232: ...APPENDIX 10 Appendices ...
Страница 237: ......
Страница 238: ......