APPENDIX 4
―――――――――――――――――――――――――――――――――――――――
Appendices
―――――――――――――――――――――――――――――――――――
The frequency for each range, when 25 samples per
period is taken as the limit.
TIME/DIV
Sampling period
Measurement
limit frequency
200
μ
s/DIV
2.5
μ
s
16 kHz
400
μ
s/DIV
5
μ
s
8 kHz
1 ms/DIV
12.5
μ
s
3.2 kHz
2 ms/DIV
25
μ
s
1.6 kHz
5 ms/DIV
62.5
μ
s
640 Hz
10 ms/DIV
125
μ
s
320 Hz
20 ms/DIV
250
μ
s
160 Hz
50 ms/DIV
625
μ
s
64 Hz
100 ms/DIV
1.25 ms
32 Hz
200 ms/DIV
2.5 ms
16 Hz
500 ms/DIV
6.25 ms
6.4 Hz
1 s/DIV
12.5 ms
3.2 Hz
2 s/DIV
25 ms
1.6 Hz
5 s/DIV
62.5 ms
0.64 Hz
10 s/DIV
125 ms
0.32 Hz
30 s/DIV
375 ms
0.10 Hz
1 min/DIV
750 ms
0.05 Hz
2 min/DIV
1.5 s
0.025 Hz