91
―――――――――――――――――――――――――――――――――――――――
7.3 Basic Operational Procedures
―――――――――――――――――――――――――――――――――――
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
APP
IND
7.3.2 Example of Operation
This example illustrates the procedure connecting the
8804 to an oscillator to measure and record a 8 Vp-p
1 Hz sine wave input.
(1) Turn on the power.
Turn on the power switch of the 8804.
(2) Connect the input.
Connect the oscillator to the input terminal of channel 1
of the 8804.
Set the oscillator so that it outputs a sine wave of
frequency 1 Hz and voltage of 8 Vp-p.
(3) Set the status screen.
1. Using the
CURSOR
keys and
+
,
−
keys, set as same as the setting on
the figure left.
(4) Set the trigger screen.
1. Press the
MENU/WAVE
key to display
the trigger screen.
2. Using the
CURSOR
keys and
+
,
−
keys, set as same as the setting on
the figure left.